TrueForce: a new digital manometer to measure maximal respiratory pressures at functional residual capacity

被引:2
|
作者
Pereira M.C.B. [1 ]
Silveira B.M.F. [2 ]
Pereira H.L.A. [2 ]
Parreira V.F. [3 ]
Martins H.R. [4 ]
机构
[1] Electrical Engineering Program, Universidade Federal de Minas Gerais, Belo Horizonte
[2] Rehabilitation Sciences Program, Universidade Federal de Minas Gerais, Belo Horizonte
[3] Department of Physiotherapy, Universidade Federal de Minas Gerais, Belo Horizonte
[4] Department of Electrical Engineering, Engineering School, Universidade Federal de Minas Gerais, Avenida Antônio Carlos, 6627, Pampulha, Belo Horizonte, 31270-901, MG
关键词
Concurrent validity; Embedded system; Manovacuometry; Respiratory muscle strength; Test-retest reliability;
D O I
10.1007/s42600-021-00139-2
中图分类号
学科分类号
摘要
Purpose: This study aimed to develop a digital manometer capable of measuring the maximal respiratory pressures at functional residual capacity level—pulmonary volume where the pressures related to the elastic recoil of the respiratory system are minimized—and to validate its functionally in healthy individuals. Methods: The proposed device was built based on three main components: the hardware, electronic part; the firmware, computational part; and the mechanics, composed of the device housing, connections, and filters used. Also, the software was updated to allow measurements at functional residual capacity. Functional validation was performed with 30 healthy individuals (15 men), aged between 20 and 40 years. The intraclass correlation coefficient (ICC) verified concurrent validity and test-retest reliability reproducibility, and pressures at functional residual capacity were compared to pressures at residual volume and total lung pulmonary capacity, by paired t test. Results: The developed device met the desired functional and non-functional requirements. The volume calculated from the acquired flow values showed variability of less than 2%. There was total agreement between the expected and obtained pressures. Regarding functional validation, an ICC ≥ 0.84; p < 0.001 was observed for both concurrent validity and test-retest reliability. Measurements at functional residual capacity were significantly lower than measurements performed at residual volume and total lung capacity (p < 0.05). Conclusion: The TrueForce can be considered a valid and reliable instrument for measures of maximum respiratory pressures performed in the functional residual capacity, and it can be used in clinical trial research. © 2021, Sociedade Brasileira de Engenharia Biomedica.
引用
收藏
页码:181 / 191
页数:10
相关论文
共 50 条
  • [41] Functional residual capacity and respiratory mechanics as indicators of aeration and collapse in experimental lung injury
    Rylander, C
    Högman, M
    Perchiazzi, G
    Magnusson, A
    Hedenstierna, G
    ANESTHESIA AND ANALGESIA, 2004, 98 (03): : 782 - 789
  • [42] RESPIRATORY MECHANICS NEAR FUNCTIONAL RESIDUAL CAPACITY IN CHRONIC OBSTRUCTIVE PULMONARY-DISEASE
    GARRITY, ER
    DRUZ, WS
    CARO, ES
    SHARP, JT
    AMERICAN REVIEW OF RESPIRATORY DISEASE, 1986, 133 (04): : A44 - A44
  • [43] MAXIMAL EXPIRATORY FLOWS AT FUNCTIONAL RESIDUAL CAPACITY - TEST OF LUNG-FUNCTION FOR YOUNG-CHILDREN
    TAUSSIG, LM
    AMERICAN REVIEW OF RESPIRATORY DISEASE, 1977, 116 (06): : 1031 - 1038
  • [44] MEASUREMENTS OF FUNCTIONAL RESIDUAL CAPACITY (FRC) AND MAXIMAL FLOW AT FRC (VMAXFRC) AS DISCRIMINATORS OF NORMAL VS ABNORMAL
    WALL, MA
    MISLEY, MC
    DICKERSON, D
    AMERICAN REVIEW OF RESPIRATORY DISEASE, 1983, 127 (04): : 223 - 223
  • [45] FUNCTIONAL RESIDUAL CAPACITY AND VENTILATORY PRESSURES DURING POSITIVE-PRESSURE VENTILATION AT HIGH-FREQUENCIES
    JONZON, A
    RONDIO, Z
    SEDIN, G
    BRITISH JOURNAL OF ANAESTHESIA, 1980, 52 (04) : 395 - 402
  • [46] TAKING THE MEASURE OF VOCATIONAL LOSS - THE USE OF RESIDUAL FUNCTIONAL-CAPACITY FORMS IN DISABILITY EVALUATION
    GRIFFIN, RE
    TRIAL, 1986, 22 (05): : 79 - 82
  • [47] NEW APPROACH FOR MEASURING FUNCTIONAL RESIDUAL CAPACITY (FRC) IN INTUBATED INFANT
    FOX, WW
    SCHWARTZ, JG
    SHAFFER, TH
    PEDIATRIC RESEARCH, 1977, 11 (04) : 570 - 570
  • [48] Functional residual capacity and compliance of the respiratory system after surfactant treatment in premature infants with severe respiratory distress syndrome
    Dinger, J
    Töpfer, A
    Schaller, P
    Schwarze, R
    EUROPEAN JOURNAL OF PEDIATRICS, 2002, 161 (09) : 485 - 490
  • [49] Functional residual capacity and compliance of the respiratory system after surfactant treatment in premature infants with severe respiratory distress syndrome
    Jürgen Dinger
    Andreas Töpfer
    Peter Schaller
    Roland Schwarze
    European Journal of Pediatrics, 2002, 161 : 485 - 490
  • [50] Impact of nursing care on lung functional residual capacity in acute respiratory distress syndrome patients
    Burey, Julien
    Guitard, Pierre-Gildas
    Girard, Noemie
    Cassiau, Florence
    Veber, Benoit
    Clavier, Thomas
    NURSING IN CRITICAL CARE, 2022, 27 (05) : 652 - 657