TrueForce: a new digital manometer to measure maximal respiratory pressures at functional residual capacity

被引:2
|
作者
Pereira M.C.B. [1 ]
Silveira B.M.F. [2 ]
Pereira H.L.A. [2 ]
Parreira V.F. [3 ]
Martins H.R. [4 ]
机构
[1] Electrical Engineering Program, Universidade Federal de Minas Gerais, Belo Horizonte
[2] Rehabilitation Sciences Program, Universidade Federal de Minas Gerais, Belo Horizonte
[3] Department of Physiotherapy, Universidade Federal de Minas Gerais, Belo Horizonte
[4] Department of Electrical Engineering, Engineering School, Universidade Federal de Minas Gerais, Avenida Antônio Carlos, 6627, Pampulha, Belo Horizonte, 31270-901, MG
关键词
Concurrent validity; Embedded system; Manovacuometry; Respiratory muscle strength; Test-retest reliability;
D O I
10.1007/s42600-021-00139-2
中图分类号
学科分类号
摘要
Purpose: This study aimed to develop a digital manometer capable of measuring the maximal respiratory pressures at functional residual capacity level—pulmonary volume where the pressures related to the elastic recoil of the respiratory system are minimized—and to validate its functionally in healthy individuals. Methods: The proposed device was built based on three main components: the hardware, electronic part; the firmware, computational part; and the mechanics, composed of the device housing, connections, and filters used. Also, the software was updated to allow measurements at functional residual capacity. Functional validation was performed with 30 healthy individuals (15 men), aged between 20 and 40 years. The intraclass correlation coefficient (ICC) verified concurrent validity and test-retest reliability reproducibility, and pressures at functional residual capacity were compared to pressures at residual volume and total lung pulmonary capacity, by paired t test. Results: The developed device met the desired functional and non-functional requirements. The volume calculated from the acquired flow values showed variability of less than 2%. There was total agreement between the expected and obtained pressures. Regarding functional validation, an ICC ≥ 0.84; p < 0.001 was observed for both concurrent validity and test-retest reliability. Measurements at functional residual capacity were significantly lower than measurements performed at residual volume and total lung capacity (p < 0.05). Conclusion: The TrueForce can be considered a valid and reliable instrument for measures of maximum respiratory pressures performed in the functional residual capacity, and it can be used in clinical trial research. © 2021, Sociedade Brasileira de Engenharia Biomedica.
引用
收藏
页码:181 / 191
页数:10
相关论文
共 50 条
  • [21] RESPIRATORY COMPLIANCE AND FUNCTIONAL RESIDUAL CAPACITY IN YOUNG-CHILDREN
    GREENOUGH, A
    STOCKS, J
    NOTHEN, U
    HELMS, P
    LOFTRUS, J
    PRICE, J
    THORAX, 1986, 41 (03) : 226 - 226
  • [22] A COMPUTERIZED TECHNIQUE TO MEASURE FUNCTIONAL RESIDUAL CAPACITY IN PATIENTS ON MECHANICAL VENTILATION
    CARLON, GC
    MIODOWNIK, S
    RAY, C
    GUY, Y
    GROEGER, JS
    CRITICAL CARE MEDICINE, 1984, 12 (03) : 274 - 274
  • [23] New reference equations for maximal respiratory pressures and maximal voluntary ventilation in the Brazilian population: a multicenter study
    Araujo, Palomma Saldanha
    Resqueti, Vanessa
    Silva, Ester
    Moreno, Marlene Aparecida
    de Andrade, Armele Dornelas
    Fregonezi, Guilehre
    EUROPEAN RESPIRATORY JOURNAL, 2017, 50
  • [24] TOTAL RESPIRATORY COMPLIANCE AND FUNCTIONAL RESIDUAL CAPACITY IN YOUNG-CHILDREN
    GREENOUGH, A
    STOCKS, J
    NOTHEN, U
    HELMS, P
    PEDIATRIC PULMONOLOGY, 1986, 2 (06) : 321 - 326
  • [25] FUNCTIONAL RESIDUAL CAPACITY AND SEVERITY OF RESPIRATORY-DISTRESS SYNDROME IN INFANTS
    RICHARDSON, P
    WYMAN, ML
    JUNG, AL
    CRITICAL CARE MEDICINE, 1980, 8 (11) : 637 - 640
  • [26] FUNCTIONAL RESIDUAL CAPACITY AND TOTAL RESPIRATORY SYSTEM IMPEDANCE IN WHEEZING INFANTS
    DESAGER, KN
    VANBEVER, HP
    WILLEMEN, M
    DEBACKER, W
    VERMEIRE, PA
    PEDIATRIC PULMONOLOGY, 1994, 17 (06) : 354 - 358
  • [27] REFERENCE VALUES FOR FUNCTIONAL RESIDUAL CAPACITY AND MAXIMAL EXPIRATORY FLOW IN YOUNG-CHILDREN
    BUIST, AS
    ADAMS, BE
    SEXTON, GJ
    AZZAM, AH
    AMERICAN REVIEW OF RESPIRATORY DISEASE, 1980, 122 (06): : 983 - 988
  • [28] In vitro validation of an ultrasonic flowmeter in order to measure the functional residual capacity in newborns
    Wauer, J
    Leier, TU
    Henschen, M
    Wauer, RR
    Schmalisch, G
    PHYSIOLOGICAL MEASUREMENT, 2003, 24 (02) : 355 - 365
  • [29] Apnea index, aerobic capacity, and maximal respiratory pressures in untreated sleep apnea patients. Colombia
    Motta, Julian Andres Rivera
    Vergara, John James Castillo
    Carrillo, Mauricio Hernandez
    Mejia, Lina Marcela Orozco
    EUROPEAN RESPIRATORY JOURNAL, 2024, 64
  • [30] RESPIRATORY MECHANICS AND FUNCTIONAL RESIDUAL CAPACITY DURING HIGH-FREQUENCY VENTILATION
    WOOD, B
    KAMA, P
    ADAMS, A
    STENZLER, A
    PEDIATRIC RESEARCH, 1994, 35 (04) : A358 - A358