Prompt gamma neutron activation analysis of boron with a 241Am-Be neutron source

被引:0
|
作者
Ş. Turhan
H. Yücel
A. Demirbaş
机构
[1] Ankara Nuclear Research and Training Center (ANRTC),
关键词
Boron; Boric Acid; Neutron Source; Boron Content; Fluence Rate;
D O I
10.1007/s10967-005-0489-z
中图分类号
学科分类号
摘要
A prompt gamma neutron activation analysis (PGAA) setup installed at ANRTC has been used to analyze boron. It consists of a 22.6% REGe detector and a 740 GBq 241Am-Be neutron source moderated with water and paraffin. At the sample irradiation position, the thermal neutron fluence rate measured was 2.36·104 n·m−2· s−1 and the corresponding Cd-ratio was 22 for gold monitor. The absolute detection efficiency in the range of 120–1500 keV was determined using 152Eu standard solution. The sensitivity and detection limit for standard boric acid samples has been determined. The boron content in boric acid prepared from Turkish borate ores is measured to be 15.91±0.46% wt.
引用
收藏
页码:661 / 664
页数:3
相关论文
共 50 条
  • [41] Prompt gamma-ray analysis using cold and thermal guided neutron beams at JAERI
    Chushiro Yonezawa
    Biological Trace Element Research, 1999, 71-72 : 407 - 413
  • [42] Introduction of the prompt γ-ray neutron activation analysis system at CARR and the first pilot experiment on boron-containing high-temperature alloys
    Tang, Chanjuan
    Yao, Yonggang
    Xiao, Caijin
    Zhao, Shouzhi
    APPLIED RADIATION AND ISOTOPES, 2024, 212
  • [43] Neutron beam preparation with Am-Be source for analysis of biological samples with PGNAA method
    Ghorbani, P.
    Sardari, D.
    Bayat, E.
    Doostmohammadi, V.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2012, 291 (03) : 839 - 842
  • [44] Progress of prompt gamma activation analysis in Korea
    C. S. Park
    G. M. Sun
    S. H. Byun
    H. D. Choi
    Journal of Radioanalytical and Nuclear Chemistry, 2005, 265 : 283 - 286
  • [45] Progress of prompt gamma activation analysis in Korea
    Park, CS
    Sun, GM
    Byun, SH
    Choi, HD
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2005, 265 (02) : 283 - 286
  • [46] Energy measurement of 241Am-9Be neutrons by tagged neutron time-of-flight
    Shana, F. S.
    Golda, K. S.
    Saneesh, N.
    Arora, D.
    Sugathan, P.
    APPLIED RADIATION AND ISOTOPES, 2023, 193
  • [47] Boron concentration measurements by prompt gamma activation analysis: Application on miocene-quaternary volcanics of the Carpathian-Pannonian Region
    K. Gméling
    Zs. Kasztovszky
    L. Szentmiklósi
    Zs. Révay
    Sz. Harangi
    Journal of Radioanalytical and Nuclear Chemistry, 2007, 271 : 397 - 403
  • [48] Conceptual study of a compact accelerator-driven neutron source for radioisotope production, boron neutron capture therapy and fast neutron therapy
    Angelone, M
    Atzeni, S
    Rollet, S
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 487 (03): : 585 - 594
  • [49] Development of high flux thermal neutron generator for neutron activation analysis
    Vainionpaa, Jaakko H.
    Chen, Allan X.
    Piestrup, Melvin A.
    Gary, Charles K.
    Jones, Glenn
    Pantell, Richard H.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2015, 350 : 88 - 93
  • [50] Instrumental neutron absorption activation analysis
    Steinhauser, Georg
    Merz, Stefan
    Sterba, Johannes H.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2013, 296 (01) : 165 - 168