Cobalt oxide nanostructure-modified glassy carbon electrode as a highly sensitive flow injection amperometric sensor for the picomolar detection of insulin

被引:0
作者
Abdollah Salimi
Rahman Hallaj
机构
[1] University of Kurdistan,Department of Chemistry
[2] University of Kurdistan,Research Center for Nanotechnology
来源
Journal of Solid State Electrochemistry | 2012年 / 16卷
关键词
Cobalt oxide; Nanostructures; Electrode; Insulin; Amperometry; Flow injection;
D O I
暂无
中图分类号
学科分类号
摘要
Glassy carbon electrode modified with electrodeposited cobalt oxide nanostructure shows an excellent electrocatalytic activity toward insulin oxidation at a wide pH range. Cyclic voltammetry, hydrodynamic amperometry, and flow injection analysis (FIA) were used for insulin determination at a picomolar and higher-concentration range. Amperometric determination of insulin at this modified electrode yielded a calibration curve with the following characteristics; linear range, 100 pM–15 nM; sensitivity of 83.9 nA nM−1 and detection limit 10 pM. FIA yielded the calibration curve with sensitivity and detection limit of 2.0 nA nM−1 and 25 pM, respectively. Furthermore, the RSD of repetitive FIA for 200 pM insulin (n = 13) is 2%. In addition, the interference effect of electroactive existing species (lactic acid, cholesterol, ascorbic acid, uric acid, and glucose) was eliminated by covering the surface of the modified electrode with nafion film. Fast response time, signal stability, high sensitivity, low cost, and ease of preparation are the advantages of the proposed insulin sensor.
引用
收藏
页码:1239 / 1246
页数:7
相关论文
共 239 条
  • [21] Kim MG(2004)undefined Analyst 129 9-14
  • [22] Shin TJ(2007)undefined Anal Chim Acta 594 24-31
  • [23] Shin HJ(2009)undefined Electrochim Acta 54 6166-6171
  • [24] Cho J(2002)undefined J Electroanal Chem 534 31-38
  • [25] Wu RJ(2003)undefined Electrochim Acta 48 3423-3429
  • [26] Wu JG(2008)undefined Sens Actuators B 129 246-254
  • [27] Tsai TK(2007)undefined Electrochim Acta 52 3654-3657
  • [28] Yeh CT(2007)undefined Biophy Chem 130 122-131
  • [29] Tabata S(2008)undefined Sens Actuators B 135 304-308
  • [30] Higaki K(2000)undefined Diabetes 76 2094-2101