Signal encoding in magnetic particle imaging: Properties of the system function

被引:246
作者
Rahmer J. [1 ]
Weizenecker J. [1 ]
Gleich B. [1 ]
Borgert J. [1 ]
机构
[1] Philips Research Europe - Hamburg, 22335 Hamburg
关键词
System Function; Magnetization Curve; Chebyshev Polynomial; Ideal Particle; Drive Frequency;
D O I
10.1186/1471-2342-9-4
中图分类号
学科分类号
摘要
Background: Magnetic particle imaging (MPI) is a new tomographic imaging technique capable of imaging magnetic tracer material at high temporal and spatial resolution. Image reconstruction requires solving a system of linear equations, which is characterized by a "system function" that establishes the relation between spatial tracer position and frequency response. This paper for the first time reports on the structure and properties of the MPI system function. Methods: An analytical derivation of the 1D MPI system function exhibits its explicit dependence on encoding field parameters and tracer properties. Simulations are used to derive properties of the 2D and 3D system function. Results: It is found that for ideal tracer particles in a harmonic excitation field and constant selection field gradient, the 1D system function can be represented by Chebyshev polynomials of the second kind. Exact 1D image reconstruction can thus be performed using the Chebyshev transform. More realistic particle magnetization curves can be treated as a convolution of the derivative of the magnetization curve with the Chebyshev functions. For 2D and 3D imaging, it is found that Lissajous excitation trajectories lead to system functions that are closely related to tensor products of Chebyshev functions. Conclusion: Since to date, the MPI system function has to be measured in time-consuming calibration scans, the additional information derived here can be used to reduce the amount of information to be acquired experimentally and can hence speed up system function acquisition. Furthermore, redundancies found in the system function can be removed to arrive at sparser representations that reduce memory load and allow faster image reconstruction. © 2009 Rahmer et al; licensee BioMed Central Ltd.
引用
收藏
相关论文
共 14 条
[1]  
Gleich B., Weizenecker J., Tomographic imaging using the nonlinear response of magnetic particles, Nature, 435, pp. 1214-1217, (2005)
[2]  
Gleich B., Weizenecker J., Borgert J., Experimental results on fast 2D-encoded magnetic particle imaging, Phys Med Biol, 53, (2008)
[3]  
Weizenecker J., Borgert J., Gleich B., A simulation study on the resolution and sensitivity of magnetic particle imaging, Phys Med Biol, 52, pp. 6363-6374, (2007)
[4]  
Weizenecker J., Gleich B., Rahmer J., Dahnke H., Borgert J., Three-dimensional real-time in vivo magnetic particle imaging, Phys Med Biol, 54, (2009)
[5]  
Liang Z.P., Lauterbur P.C., Principles of Magnetic Resonance Imaging, (2000)
[6]  
Chikazumi S., Charap S.H., Physics of Magnetism, (1964)
[7]  
Bean C.P., Livingston J.D., Superparamagnetism, J Appl Phys, 30, (1959)
[8]  
Weizenecker J., Gleich B., Borgert J., Magnetic particle imaging using a field free line, J Phys D: Appl Phys, 41, (2008)
[9]  
Boyd J.P., Chebyshev and Fourier Spectral Methods, (2001)
[10]  
Press W.H., Teukolsky S.A., Vetterling W.T., Flannery B.P., Numerical Recipes in C, (1992)