共 50 条
Development and Characterization of a Portable Total Reflection X-ray Fluorescence System Using a Waveguide for Trace Elements Analysis
被引:0
|作者:
Ana Cristina M. da Costa
Ubiratan B. da Araújo
Edgar F. O. da Jesus
Marcelino J. Anjos
Ricardo T. Lopes
机构:
[1] Nuclear Instrumentation Laboratory–COPPE/UFRJ,
[2] State University of Rio de Janeiro,undefined
[3] Physics Institute,undefined
来源:
Analytical Sciences
|
2014年
/
30卷
关键词:
X-ray fluorescence;
total-reflection;
trace elements;
waveguide;
D O I:
暂无
中图分类号:
学科分类号:
摘要:
This paper presents a portable total reflection X-ray fluorescence system composed of a 15 W X-ray tube, with a gold anode, a waveguide constituted by two Perspex® parallel plates, a Si PIN detector and a quartz optical flat. The critical angle of the total reflection system was experimentally determined by measuring a zinc solution (100 mg/L). The accuracy of the system was checked using SRM 1577b Bovine Liver by NIST as standard reference material. We obtained the absolute detection limits of the following elements: P (450 ± 40 ng), S (200 ± 31 ng), K (30 ± 2.5 ng), Ca (19 ± 3.5 ng), Mn (4.1 ± 0.5 ng), Fe (3.6 ± 0.9 ng), Cu (3.3 ± 0.4 ng) and Zn (3.5 ± 0.3 ng). This paper shows that it is possible to produce total reflection X-ray fluorescence with very compact, efficient, low-cost and easy-to-handle instrumentation using a low-power X-ray tube and a Si PIN compact detector.
引用
收藏
页码:955 / 960
页数:5
相关论文