Corrosion Behavior of Selective Laser Melting (SLM) Manufactured Ti6Al4V Alloy in Saline and BCS Solution

被引:0
作者
Fischer D. [1 ]
Cheng K.-Y. [1 ]
Neto M.Q. [2 ]
Hall D. [2 ]
Bijukumar D. [1 ]
Espinoza Orías A.A. [2 ]
Pourzal R. [2 ]
van Arkel R.J. [3 ]
Mathew M.T. [1 ]
机构
[1] Department of Biomedical Science, University of Illinois College of Medicine Rockford, Rockford, IL
[2] Department of Orthopedic Surgery, Rush University Medical Center, Chicago, IL
[3] Department of Mechanical Engineering, Imperial College, London
基金
美国国家卫生研究院;
关键词
Additive manufacturing; Corrosion; Implant medicine; Selective laser melting (SLM); Ti6Al4V;
D O I
10.1007/s40735-022-00657-1
中图分类号
学科分类号
摘要
The frequency of surgeries involving the use of metal implants in orthopedic medicine to replace degenerative or fractured joints is increasing, and it is therefore important to optimize the lifespan and quality of these implants. Advances in additive manufacturing (AM), or 3D printing, are creating new opportunities to personalize implants in ways that reduce mechanical stress at the joint implant interface and improve bone ingrowth and implant stability; however, it is not well understood if and to what degree the AM process alters the corrosion behavior of the materials it produces. In this study, six Ti6Al4V prints manufactured via a selective laser melting (SLM) method were examined regarding their corrosion behavior in both saline and bovine calf serum (BCS) solutions. Ecorr and Icorr values were comparable between the CM-Ti6Al4V control and SLM-EDM surfaces; however, SLM surfaces were found to have more narrow passivation behavior evidenced by significant decreases in Epass values relative to CM-Ti6Al4V. We believe this is a consequence of microstructural differences between CM-Ti6Al4V and SLM-Ti6Al4V. Specifically, the SLM-Ti6Al4V demonstrated a dominant α′ martensitic microstructure and decreased vanadium-rich β-phase. BCS solution had a detrimental effect on potential parameters, Ecorr and OCP, decreasing these values relative to their saline counterparts. Increased surface roughness of the SLM-printed surface seemed to amplify the effects of the BCS solution. Furthermore, modest decreases in Epass and Ipass were observed in BCS solution, suggesting that the presence of protein may also interfere with passivation behavior. These findings have implications for how SLM-Ti6Al4V implants will perform in vivo and could possibly influence implant longevity and performance. © 2022, The Author(s), under exclusive licence to Springer Nature Switzerland AG.
引用
收藏
相关论文
共 35 条
[1]  
Kurtz S.M., Ong K.L., Lau E., Bozic K.J., Impact of the economic downturn on total joint replacement demand in the United States: updated projections to 2021, J Bone Joint Surg Am, 96, pp. 624-630, (2014)
[2]  
Ravi B., Et al., The changing demographics of total joint arthroplasty recipients in the United States and Ontario from 2001 to 2007, Best Pract Res Clin Rheumatol, 26, pp. 637-647, (2012)
[3]  
Bayliss L.E., Et al., The effect of patient age at intervention on risk of implant revision after total replacement of the hip or knee: a population-based cohort study, The Lancet, 389, pp. 1424-1430, (2017)
[4]  
Huang H.-H., Effect of fluoride and albumin concentration on the corrosion behavior of Ti–6Al–4V alloy, Biomaterials, 24, pp. 275-282, (2003)
[5]  
Guillemot F., Recent advances in the design of titanium alloys for orthopedic applications, Expert Rev Med Devices, 2, pp. 741-748, (2005)
[6]  
Rack H.J., Qazi J.I., Titanium alloys for biomedical applications, Mater Sci Eng C, 26, pp. 1269-1277, (2006)
[7]  
Geetha M., Singh A.K., Asokamani R., Gogia A.K., Ti based biomaterials, the ultimate choice for orthopaedic implants—a review, Prog Mater Sci, 54, pp. 397-425, (2009)
[8]  
Srivastav A., An overview of metallic biomaterials for bone support and replacement, Biomed Eng Trends Mater Sci, (2011)
[9]  
Bagaria V., Bhansali R., Pawar P., 3D printing-creating a blueprint for the future of orthopedics: current concept review and the road ahead!, Journal of Clinical Orthopaedics and Trauma, 9, pp. 207-212, (2018)
[10]  
Sidambe A.T., Biocompatibility of advanced manufactured titanium implants—a review, Materials, 7, pp. 8168-8188, (2014)