Study on simulation and experiment of array micro Faraday cup ion detector for FAIMS

被引:0
作者
Fei Tang
XiaoHao Wang
Liang Zhang
ZiLin Yan
机构
[1] Tsinghua University,State Key Laboratory of Precision Measurement Technology and Instruments, Department of Precision Instruments and Mechanology
来源
Science China Technological Sciences | 2010年 / 53卷
关键词
FAIMS; MEMS; Faraday cup; ion detector; low current;
D O I
暂无
中图分类号
学科分类号
摘要
An array micro Faraday cup ion detector for FAIMS (High-field Asymmetric Waveform Ion Mobility Spectrometry) was introduced, with which the size of the FAIMS system was reduced. With simple structure, good stability, low noise, large measurements range, this detector can work under the condition of atmospheric pressure. The array micro Faraday cup was composed of gate electrode, sensitive electrode and shielding electrode. The sensitive electrode was made of tens of crossing silicon columns with diameter of 200 μm. It was fabricated through typical MEMS technology, which was compatible completely with plane FAIMS. It was shown from FLUENT simulation result that the resistance to gas motion was low and the flow field distribution was helpful for full absorption of the ion in this array design. Through electricity simulation, the method to reduce interference on the detection signal of the micro Faraday cup was given out. Connecting with KEITHLEY 237 ampere meter, the noise level of the array micro Faraday cup was lower than 0.5 pA. The output signal of the acetone sample measured by experiment was about 210 pA. Through contrast experiment, the design parameter of the micro Faraday cup was obtained primarily. This array micro Faraday cup can meet the requirements of the FAIMS system.
引用
收藏
页码:3225 / 3231
页数:6
相关论文
共 39 条
[1]  
Miller R. A.(2000)A novel micro-machined high-field asymmetric waveform-ion mobility spectrometer Sensors and Actuators B 67 300-306
[2]  
Eiceman G. A.(2004)High-field asymmetric waveform ion mobility spectrometry: A new tool for mass spectrometry J Chromatography A 1058 3-19
[3]  
Nazarov E. G.(2002)CCD readout of GEM-based neutron detectors Nucl Instrum Meth Phys Res A 478 357-361
[4]  
Guevremont R.(2002)Measurement of low currents in an external beam set-up Nucl Instrum Meth Phys Res B 188 162-165
[5]  
Fraga F. A. F.(2007)Experimental study on high-voltage solar array sustained arc discharge induced by high charging Sci China Ser E-Tech Sci 50 213-220
[6]  
Margato L. M. S.(2008)A new approach to solar wind monitoring Adv Space Res 41 153-159
[7]  
Fetal S. T. G.(2005)Comparison of two particle charging experiments for testing the JSC Mars-1 Martian regolith simulant against man-made materials IEEE Trans Dielec Elec Insul 12 821-826
[8]  
Chiari M.(2002)Micromachined mass spectrometer Sensors and Actuators B 83 129-137
[9]  
Migliori A.(2003)Analysis of the dynamic Faraday cup J Phys D: Appl Phys 36 2407-2416
[10]  
Mandò P. A.(2007)Study on interactions of He atoms and ions with an Al(1 1 1) surface Nucl Instrum Meth Phys Res B 258 7-12