Rapid lipid bilayer membrane formation on Parylene coated apertures to perform ion channel analyses

被引:0
|
作者
Tanzir Ahmed
Sander van den Driesche
Jayesh Arun Bafna
Martin Oellers
Roland Hemmler
Karsten Gall
Richard Wagner
Mathias Winterhalter
Michael J. Vellekoop
机构
[1] University of Bremen,Institute for Microsensors,
[2] Microsystems Center Bremen (MCB),actuators and
[3] Jacobs University Bremen,systems (IMSAS)
[4] Ionovation GmbH,Department of Life Sciences and Chemistry
来源
Biomedical Microdevices | 2020年 / 22卷
关键词
Parylene-C; Parylene-AF4; Lipid Bilayer Membrane (LBL); Silicon; Silicon Nitride;
D O I
暂无
中图分类号
学科分类号
摘要
We present a chip design allowing rapid and robust lipid bilayer (LBL) membrane formation using a Parylene coated thin silicon nitride aperture. After bilayer formation, single membrane channels can be reconstituted and characterized by electrophysiology. The ability for robust reconstitution will allow parallelization and enhanced screening of small molecule drugs acting on or permeating across the membrane channel. The aperture was realized on a microfabricated silicon nitride membrane by using standard clean-room fabrication processes. To ensure the lipid bilayer formation, the nitride membrane was coated with a hydrophobic and biocompatible Parylene layer. We tested both Parylene-C and Parylene-AF4. The contact angle measurements on both Parylene types showed very good hydrophobic properties and affinity to lipids. No precoating of the Parylene with an organic solvent is needed to make the aperture lipophilic, in contradiction to Teflon membranes. The chips can be easily placed in an array utilizing a 3D printed platform. Experiments show repetitive LBL formation and destruction (more than 6 times) within a very short time (few seconds). Through measurements we have established that the LBL layers are very thin. This allows the investigation of the fusion process of membrane proteins i.e. outer membrane protein (OmpF) in the LBL within a few minutes.
引用
收藏
相关论文
共 50 条
  • [41] Voltage-induced gating of the mechanosensitive MscL ion channel reconstituted in a tethered lipid bilayer membrane
    Andersson, Martin
    Okeyo, George
    Wilson, Danyell
    Keizer, Henk
    Moe, Paul
    Blount, Paul
    Fine, Daniel
    Dodabalapur, Ananth
    Duran, Randolph S.
    BIOSENSORS & BIOELECTRONICS, 2008, 23 (06): : 919 - 923
  • [42] Role of polyol moiety of amphotericin B in ion channel formation and sterol selectivity in bilayer membrane
    Yamamoto, Tomoya
    Umegawa, Yuichi
    Tsuchikawa, Hiroshi
    Matsumori, Nobuaki
    Hanashima, Shinya
    Murata, Michio
    Haser, Resul
    Rawlings, Bernard J.
    Caffrey, Patrick
    BIOORGANIC & MEDICINAL CHEMISTRY, 2015, 23 (17) : 5782 - 5788
  • [43] Lipid bilayer-coated silica beads in membrane fusion assays
    Verbeek, S. F.
    Savic, F.
    Ortega, R. D. Siaden
    Kliesch, T. -T.
    Janshoff, A.
    EUROPEAN BIOPHYSICS JOURNAL WITH BIOPHYSICS LETTERS, 2017, 46 : S129 - S129
  • [44] Lipid bilayer technologies for high throughput ion channel screening
    Schmidt, Jacob
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2009, 237
  • [45] Artificial lipid bilayer platforms for human ion channel screening
    Schmidt, Jacob
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2011, 241
  • [46] Lipid bilayer array for simultaneous recording of ion channel activities
    Hirano-Iwata, Ayumi
    Nasu, Tomohiro
    Oshima, Azusa
    Kimura, Yasuo
    Niwano, Michio
    APPLIED PHYSICS LETTERS, 2012, 101 (02)
  • [47] CONFORMATION OF SEQUENTIAL POLYPEPTIDE POLY(LEU-LEU-D-PHE-PRO) AND FORMATION OF ION CHANNEL ACROSS BILAYER LIPID-MEMBRANE
    KAMEGAI, J
    KIMURA, S
    IMANISHI, Y
    BIOPHYSICAL JOURNAL, 1986, 49 (05) : 1101 - 1108
  • [48] Protein-fluctuation-induced water-pore formation in ion channel voltage-sensor translocation across a lipid bilayer membrane
    Rajapaksha, Suneth P.
    Pal, Nibedita
    Zheng, Desheng
    Lu, H. Peter
    PHYSICAL REVIEW E, 2015, 92 (05):
  • [49] Membrane Activity and Channel Formation of the Adenylate Cyclase Toxin (CyaA) of Bordetella pertussis in Lipid Bilayer Membranes
    Knapp, Oliver
    Benz, Roland
    TOXINS, 2020, 12 (03)
  • [50] PROPERTIES OF THE LIPID BILAYER - ION GRADIENTS, LIPID ASYMMETRY, AND MEMBRANE-FUNCTION
    CULLIS, PR
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 1995, 5 (10): : 1844 - 1844