Protein determination in serum and whole blood by attenuated total reflectance infrared spectroscopy

被引:0
|
作者
D. Perez-Guaita
J. Ventura-Gayete
C. Pérez-Rambla
M. Sancho-Andreu
S. Garrigues
M. de la Guardia
机构
[1] University of Valencia,Department of Analytical Chemistry, Research Building
[2] University Hospital Doctor Peset Aleixandre,undefined
来源
关键词
Serum protein determination; ATR-FTIR; PLS; Point-of-care; Variable subset selection; Green analytical chemistry;
D O I
暂无
中图分类号
学科分类号
摘要
Attenuated total reflectance mid-infrared spectra of serum and blood samples were obtained from 4,000 to 600 cm−1. Models for the determination of albumin, immunoglobulin, total globulin, and albumin/globulin coefficients were established for serum samples, using reference data obtained by capillary electrophoresis. Based on the use of the amide bands I and II regions, the relative root mean square error of prediction (RRMSEP) was 4.9, 14.9, 4.5, and 7.1 % for albumin, immunoglobulin, total globulin, and albumin/globulin coefficients, respectively, determined in an independent validation set of 120 samples using 200 samples for calibration. Additionally, the use of Kennard–Stone method for the selection of a representative calibration subset of samples provided comparable results using only 60 samples. For whole blood analysis, hemoglobin was determined in 40 validation samples using models built from 40 calibration independent samples with RRMSEP of 8.3, 5.5, and 4.9 % with models built from direct spectra in the first case and from sample spectra recorded after lysis by sodium dodecyl sulfate and freezing, respectively, for the last two ones. The developed methodologies offer green alternatives for patient diagnosis in a few minutes, minimizing the use of reagents and residues and being adaptable for its use as a point-of-care method.
引用
收藏
页码:649 / 656
页数:7
相关论文
共 50 条
  • [1] Protein determination in serum and whole blood by attenuated total reflectance infrared spectroscopy
    Perez-Guaita, D.
    Ventura-Gayete, J.
    Perez-Rambla, C.
    Sancho-Andreu, M.
    Garrigues, S.
    de la Guardia, M.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2012, 404 (03) : 649 - 656
  • [2] Multivariate determination of hematocrit in whole blood by attenuated total reflection infrared spectroscopy
    Kostrewa, S
    Paarmann, C
    Goemann, W
    Heise, HM
    FOURIER TRANSFORM SPECTROSCOPY, 1998, (430): : 271 - 273
  • [3] Attenuated total reflectance infrared spectroscopy
    Chaberska, Hanna
    PRZEMYSL CHEMICZNY, 2008, 87 (03): : 274 - 276
  • [4] MULTIVARIATE DETERMINATION OF GLUCOSE IN WHOLE-BLOOD BY ATTENUATED TOTAL REFLECTION INFRARED-SPECTROSCOPY
    HEISE, HM
    MARBACH, R
    JANATSCH, G
    KRUSEJARRES, JD
    ANALYTICAL CHEMISTRY, 1989, 61 (18) : 2009 - 2015
  • [5] MULTIVARIATE DETERMINATION OF SUGAR POWDERS BY ATTENUATED TOTAL REFLECTANCE INFRARED-SPECTROSCOPY
    DUPUY, N
    MEURENS, M
    SOMBRET, B
    LEGRAND, P
    HUVENNE, JP
    APPLIED SPECTROSCOPY, 1993, 47 (04) : 452 - 457
  • [6] Determination of protein concentration in raw milk by mid-infrared Fourier transform infrared/attenuated total reflectance spectroscopy
    Etzion, Y
    Linker, R
    Cogan, U
    Shmulevich, I
    JOURNAL OF DAIRY SCIENCE, 2004, 87 (09) : 2779 - 2788
  • [7] EXPERIMENTAL ASPECTS OF ATTENUATED TOTAL REFLECTANCE SPECTROSCOPY IN THE INFRARED
    OFFERMANN, V
    GROSSE, P
    FEUERBACHER, M
    DITTMAR, G
    VIBRATIONAL SPECTROSCOPY, 1995, 8 (02) : 135 - 140
  • [8] INFRARED SPECTROSCOPY OF METEORITES USING ATTENUATED TOTAL REFLECTANCE
    Reach, W. T.
    Rossman, G. R.
    METEORITICS & PLANETARY SCIENCE, 2010, 45 : A169 - A169
  • [9] Studies of adipocere using attenuated total reflectance infrared spectroscopy
    Stuart B.H.
    Craft L.
    Forbes S.L.
    Dent B.B.
    Forensic Science, Medicine, and Pathology, 2005, 1 (3) : 197 - 201
  • [10] INFRARED SPECTRA OF WOOD SURFACES BY ATTENUATED TOTAL REFLECTANCE SPECTROSCOPY
    OBRIAN, RN
    HARTMAN, K
    PULP AND PAPER MAGAZINE OF CANADA, 1969, 70 (08): : 67 - &