Measuring the actual I-131 thyroid uptake curve with a collar detector system: a feasibility study

被引:0
|
作者
Peter Brinks
Koen Van Gils
Ellen Kranenborg
Jules Lavalaye
Dennis Dieckens
Jan B. A. Habraken
机构
[1] Department of Medical Physics,
[2] St. Antonius Hospital,undefined
[3] Department of Nuclear Medicine,undefined
[4] St. Antonius Hospital,undefined
来源
European Journal of Nuclear Medicine and Molecular Imaging | 2017年 / 44卷
关键词
Thyroid; Iodine; I-131; Radionuclide therapy; Dosimetry; Pharmacokinetics;
D O I
暂无
中图分类号
学科分类号
摘要
Radionuclide therapy using I-131 is commonly used for the treatment of benign thyroid diseases. The therapeutic dose to be administered is calculated based on the type of disease, the volume of the thyroid, and the measured uptake percentage. This methodology assumes a similar biological half-life of iodine, whereas in reality a large variation in biological half-life is observed. More knowledge about the actual biological half-life of iodine for individual patients will improve the quantification of the delivered radiation dose during radioiodine therapy and could aid the evaluation of the success of the therapy. In this feasibility study we used a novel measurement device [Collar Therapy Indicator (CoTI)] to measure the uptake curve of patients undergoing I-131 radioiodine therapy. The CoTI device is a light-weight wearable device that contains two independent gamma radiation detectors that are placed in a collar. By comparing results of thyroid uptake measurements with results obtained with a gamma camera, the precision of the system is demonstrated. Additionally, for three patients the uptake curve is measured during 48 h of admission in the hospital. The presented results demonstrate the feasibility of the new measurement device to measure the uptake curve during radioiodine therapy.
引用
收藏
页码:935 / 940
页数:5
相关论文
共 7 条
  • [1] Measuring the actual I-131 thyroid uptake curve with a collar detector system: a feasibility study
    Brinks, Peter
    Van Gils, Koen
    Kranenborg, Ellen
    Lavalaye, Jules
    Dickerscheid, Dennis B. M.
    Habraken, Jan B. A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 (06) : 935 - 940
  • [2] A method to measure the absorbed dose of the thyroid during I-131 therapy, using a collar detector system and a SPECT acquisition
    van Gils, Koen
    Brinks, Peter
    Lavalaye, Jules
    Verberne, Hein J.
    Habraken, Jan B. A.
    MEDICAL PHYSICS, 2017, 44 (10) : 5450 - 5456
  • [3] Unexpected False-positive I-131 Uptake in Patients with Differentiated Thyroid Carcinoma
    Oral, Aylin
    Yazici, Bulent
    Eraslan, Cenk
    Burak, Zeynep
    MOLECULAR IMAGING AND RADIONUCLIDE THERAPY, 2018, 27 (03) : 99 - 106
  • [4] Influence of measurement geometry on the estimate of I-131 activity in the thyroid: Monte Carlo simulation of a detector and a phantom
    Ulanovsky, AV
    Minenko, VF
    Korneev, SV
    HEALTH PHYSICS, 1997, 72 (01): : 34 - 41
  • [5] POTASSIUM IODATE AND ITS COMPARISON TO POTASSIUM-IODIDE AS A BLOCKER OF I-131 UPTAKE BY THE THYROID IN RATS
    PAHUJA, DN
    RAJAN, MGR
    BORKAR, AV
    SAMUEL, AM
    HEALTH PHYSICS, 1993, 65 (05): : 545 - 549
  • [6] Clinical conditions related to positive thymic uptake on I-131 post-therapeutic scans in thyroid cancer patients
    Seo, Ye Young
    Choi, Hyun Su
    Park, Sonya Youngju
    CLINICAL ENDOCRINOLOGY, 2023, 98 (04) : 603 - 608
  • [7] Comparative study of MIRD, experimental and GEANT4 simulations for uniformly distributed I-131 in cylindrical and spherical thyroid models
    Rahman, Ziaur
    Rehman, Shakeel Ur
    Mirza, Sikander M.
    Arshed, Waheed
    Mirza, Nasir M.
    Masood, Muhammad
    RADIATION MEASUREMENTS, 2012, 47 (06) : 406 - 409