Quality control of ultrasound transducers: analysis of evaluation parameters and results of a survey of 116 transducers in a single hospital

被引:3
作者
Cozzolino, P. [1 ]
Stramare, R. [1 ]
Udilano, A. [2 ]
Castoro, M. [3 ]
Scagliori, E. [1 ]
Beltrame, V. [1 ]
Rubaltelli, L. [1 ]
机构
[1] Univ Padua, Dipartimento Sci Med Diagnost & Terapie Speciali, Padua, Italy
[2] Azienda Osped Padova, Dipartimento Ingn Clin, Padua, Italy
[3] Azienda Osped Padova, Unita Valutaz Technol Assessment, Padua, Italy
来源
RADIOLOGIA MEDICA | 2010年 / 115卷 / 04期
关键词
Ultrasound; Technology assessment; Quality control; LATERAL RESOLUTION; EQUIPMENT; PHANTOMS; SPEED; SOUND; MODE;
D O I
10.1007/s11547-010-0533-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose. The aim of this study was to develop a quality control programme for ultrasound equipment and assess equipment performance at a single hospital. Materials and methods. We tested 116 transducers by means of a phantom; ultrasound quality control tests were done on parameters from B-mode images, and each parameter was rated according to a three-point score as good, sufficient or poor. Results. Surface transducers showed worse performance in terms of image uniformity, with 7/55 (13%) transducers rated poor, and depth of penetration, with 24/55 (44%) transducers, rated sufficient or poor. The 3.5-MHz transducers showed worse performance in lateral resolution, with 18/48 (38%) rated poor, and particularly in focusing, with 40/48 (83%) classified as poor. In addition, the 3.5-MHz transducers proved less accurate in vertical distance accuracy than in horizontal distance, with 31/48 (64%) considered sufficient. No significant geometric distortions of masses were identified, although 4/116 (3%) transducers failed to visualise two masses. Conclusions. The first goal of an ultrasound quality control programme is to establish specific parameters to evaluate equipment performance and, should results be below recommended action levels, implement corrective measures.
引用
收藏
页码:668 / 677
页数:10
相关论文
共 16 条
[1]  
BAZZOCCHI M, 2002, ECOGRAFIA
[2]  
BAZZOCCHI M, 1996, 496 SIRM
[3]   Assessment of the acoustic properties of common tissue-mimicking test phantoms [J].
Browne, JE ;
Ramnarine, KV ;
Watson, AJ ;
Hoskins, PR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2003, 29 (07) :1053-1060
[4]   Simulation study of effects of speed of sound and attenuation on ultrasound lateral resolution [J].
Chen, Q ;
Zagzebski, JA .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2004, 30 (10) :1297-1306
[5]  
Doleal L, 2006, MEASUREMENT SCI REV, V6, P5
[6]   The effect of speed of sound in ultrasound test objects on lateral resolution [J].
Dudley, NJ ;
Gibson, NM ;
Fleckney, MJ ;
Clark, PD .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2002, 28 (11-12) :1561-1564
[7]   Real-time B-mode ultrasound quality control test procedures - Report of AAPM ultrasound task group no. 1 [J].
Goodsitt, MM ;
Carson, PL ;
Witt, S ;
Hykes, DL ;
Kofler, JM .
MEDICAL PHYSICS, 1998, 25 (08) :1385-1406
[8]   Implementation and validation of three automated methods for measuring ultrasound maximum depth of penetration: Application to ultrasound quality control [J].
Gorny, KR ;
Tradup, DJ ;
Hangiandreou, NJ .
MEDICAL PHYSICS, 2005, 32 (08) :2615-2628
[9]  
Gullace Giuseppe, 2002, Ital Heart J Suppl, V3, P1204
[10]  
KOFLER JM, 2001, AAPM 43 ANN M AM ASS