Correction equations of coincidence summing using75Se radionuclide in the efficiency of HpGe detector

被引:0
作者
Y. Lee
S. Park
K. Choi
T. Horiguchi
H. Inoue
机构
[1] POSTECH,Pohang Accelerator Laboratory
[2] Korea Research Institute of Standards and Science,Department of Physics, Faculty of Science
[3] Hiroshima University,undefined
来源
Journal of Radioanalytical and Nuclear Chemistry | 1999年 / 242卷
关键词
Physical Chemistry; Inorganic Chemistry; Radionuclide; Energy Region; HpGe Detector;
D O I
暂无
中图分类号
学科分类号
摘要
Correction equations of the coincidence-summing effect for efficiencies of HpGe detector based on the decay scheme were developed by considering the summing up to triple coincidence. The correction equations which do not dependent on the kind of the Ge detector are very useful for efficiency calibrations of a Ge detector in the energy region from 60 to 400 keV by using75Se radionuclide even with very short source-to-detector distances.
引用
收藏
页码:105 / 110
页数:5
相关论文
共 26 条
[1]  
Debertin K.(1979)undefined Nucl. Instr. Meth. 158 471-471
[2]  
Schotiz U.(1981)undefined Nucl. Instr. Meth. 190 101-101
[3]  
Kawade K.(1977)undefined Nucl. Instr. Meth. 147 405-405
[4]  
Ezuka M.(1980)undefined Nucl. Instr. Meth. 169 43-43
[5]  
Yamamoto H.(1972)undefined Instr. Exp. Tech. 25 1358-1358
[6]  
Sugioka K.(1975)undefined Nucl. Instr. Meth. 130 189-189
[7]  
Katoh T.(1983)undefined Intern. J. Appl. Radiation and Isotopes 34 1109-1109
[8]  
Gehrke R. J.(1983)undefined Nucl. Instr. Meth. 212 249-249
[9]  
Helmer R. G.(undefined)undefined undefined undefined undefined-undefined
[10]  
Greenwood R. C.(undefined)undefined undefined undefined undefined-undefined