Feasibility of using solid sampling graphite furnace atomic absorption spectrometry for preparation of spiked filter papers with Cu and Zn as standards for direct solid analysis

被引:0
作者
Cassiana S. Nomura
Danielle P. Intima
Pedro V. Oliveira
Iolanda A. Ruffini
Francisco J. Krug
机构
[1] Universidade Federal do ABC,Centro de Ciências Naturais e Humanas
[2] Universidade de São Paulo,Departamento de Química Fundamental, Instituto de Química
[3] Universidade de São Paulo,Centro de Energia Nuclear na Agricultura
来源
Analytical and Bioanalytical Chemistry | 2008年 / 391卷
关键词
Solid sampling; Graphite furnace; Calibrating standard; Microhomogeneity;
D O I
暂无
中图分类号
学科分类号
摘要
An approach was developed for the preparation of cryogenic ground spiked filter papers with Cu and Zn for use as synthetic calibrating standards for direct solid microanalysis. Solid sampling graphite furnace atomic absorption spectrometry was used to evaluate the microhomogeneity and to check the applicability of the synthetic calibrating standards for the direct determination of Cu and Zn in vegetable certified reference materials. The found concentrations presented no statistical differences at the 95% confidence level. The homogeneity factors ranged from 2.7 to 4.2 for Cu and from 6.4 to 11.5 for Zn.
引用
收藏
页码:1135 / 1137
页数:2
相关论文
共 43 条
[1]  
Nomura CS(2008)undefined Química Nova 31 104-113
[2]  
Oliveira PV(2006)undefined Spectrochim Acta Part B 61 454-459
[3]  
Detcheva A(2006)undefined Anal Chim Acta 571 142-149
[4]  
Grobecker KH(2007)undefined Environ Sci Technol 41 593-598
[5]  
Resano M(2006)undefined Appl Spectrosc Rev 41 131-150
[6]  
Aramendia M(2007)undefined J Braz Chem Soc 18 463-512
[7]  
Garcia-Ruiz E(1998)undefined Fresenius J Anal Chem 360 376-379
[8]  
Crespo C(1996)undefined Appl Spectrosc 50 1175-1181
[9]  
Belarra MA(2007)undefined Talanta 71 1399-1404
[10]  
Legrand M(1974)undefined Anal Chim Acta 69 267-273