Clinical evaluation of the Nanoduct sweat test system in the diagnosis of cystic fibrosis after newborn screening

被引:0
|
作者
Annette Vernooij-van Langen
Edward Dompeling
Jan-Bart Yntema
Bert Arets
Harm Tiddens
Gerard Loeber
Jeannette Dankert-Roelse
机构
[1] St Jansdal Hospital,Department of Paediatrics
[2] Maastricht University Medical Centre,Department of Paediatric Pulmonology
[3] University Medical Centre Nijmegen-St Radboud,Department of Paediatric Pulmonology
[4] University Medical Centre Utrecht-Wilhelmina’s Children’s Hospital,Department of Paediatric Pulmonology
[5] Sophia Children’s Hospital,Department of Paediatric Pulmonology, Erasmus Medical Centre
[6] National Institute for Public Health and the Environment (RIVM),Laboratory for Infectious Diseases and Perinatal Screening
[7] Atrium Medical Centre,Department of Paediatrics
来源
European Journal of Pediatrics | 2015年 / 174卷
关键词
Cystic fibrosis; Diagnosis; Infants; Sweat test; Nanoduct;
D O I
暂无
中图分类号
学科分类号
摘要
After a positive newborn screening test for cystic fibrosis (CF), a sweat test is performed to confirm the diagnosis. The success rate of the generally acknowledged methods (Macroduct/Gibson and Cooke) in newborns varies between 73 and 99 %. The Nanoduct sweat test system is easier to perform and less sweat is needed. The main aim of this study was to measure the success rate of the Nanoduct compared to current approved sweat test methods in a newborn population. After informed consent of the parents, newborns with a positive screening test for CF were included. The Macroduct or Gibson and Cooke and Nanoduct were performed in all infants, during the same appointment. The chloride concentration was determined by standard coulorimetry; conductivity was measured directly and converted to a NaCl molarity. One hundred eight newborns were included: 17 with CF, 7 with cystic fibrosis transmembrane regulator (CFTR)-related metabolic syndrome (CRMS), and 84 healthy children. The success rate of the Nanoduct was 93 % and for the Macroduct/Gibson and Cooke 79 % (McNemar, p = 0.002). The Nanoduct detected the same CF patients as the Macroduct/Gibson and Cooke; one CF patient had an equivocal result for both tests, and no patients were missed. The area under the receiver operating characteristic curve for detection of CF with the Nanoduct was 0.999, with ideal cutoff levels of 91 and 66 mmol/l, comparable to former studies.
引用
收藏
页码:1025 / 1034
页数:9
相关论文
共 50 条
  • [1] Clinical evaluation of the Nanoduct sweat test system in the diagnosis of cystic fibrosis after newborn screening
    Vernooij-van Langen, Annette
    Dompeling, Edward
    Yntema, Jan-Bart
    Arets, Bert
    Tiddens, Harm
    Loeber, Gerard
    Dankert-Roelse, Jeannette
    EUROPEAN JOURNAL OF PEDIATRICS, 2015, 174 (08) : 1025 - 1034
  • [2] Comparison of nanoduct versus macroduct sweat test for the diagnosis of cystic fibrosis in the newborn screening programme in Switzerland
    Barben, Juerg
    Rueegg, Corina S.
    Gallati, Sabina
    Kuehni, Claudia E.
    Baumgartner, Matthias
    Torresani, Toni
    Schoeni, Martin H.
    EUROPEAN RESPIRATORY JOURNAL, 2012, 40
  • [3] Parental Understanding of Newborn Screening for Cystic Fibrosis After a Negative Sweat-Test
    Lang, Colleen Walsh
    McColley, Susanna A.
    Lester, Lucille A.
    Ross, Lainie Friedman
    PEDIATRICS, 2011, 127 (02) : 276 - 283
  • [4] Sweat test practice in pediatric pulmonology after introduction of cystic fibrosis newborn screening
    Grimaldi, Celine
    Brement, Francois
    Berlioz-Baudoin, Michele
    Brouard, Jacques
    Corvol, Harriet
    Couderc, Laure
    Lezmi, Guillaume
    Pin, Isabelle
    Petit, Isabelle
    Reix, Philippe
    Remus, Natacha
    Schweitzer, Cyril
    Thumerelle, Caroline
    Dubus, Jean-Christophe
    EUROPEAN JOURNAL OF PEDIATRICS, 2015, 174 (12) : 1613 - 1620
  • [5] Nanoduct® sweat testing for rapid diagnosis in newborns, infants and children with cystic fibrosis
    Desax, Marie-Claire
    Ammann, Roland A.
    Hammer, Juerg
    Schoeni, Martin H.
    Barben, Juerg
    EUROPEAN JOURNAL OF PEDIATRICS, 2008, 167 (03) : 299 - 304
  • [6] Nanoduct® sweat testing for rapid diagnosis in newborns, infants and children with cystic fibrosis
    Marie-Claire Desax
    Roland A. Ammann
    Jürg Hammer
    Martin H. Schoeni
    Jürg Barben
    European Journal of Pediatrics, 2008, 167 : 299 - 304
  • [7] Cystic Fibrosis Diagnosis and Newborn Screening
    Rosenfeld, Margaret
    Sontag, Marci K.
    Ren, Clement L.
    PEDIATRIC CLINICS OF NORTH AMERICA, 2016, 63 (04) : 599 - +
  • [8] A sweat test centered protocol for the disclosure and diagnosis of cystic fibrosis in a newborn screening program
    Doull, I. J. M.
    Hall, S. J.
    Bradley, D. M.
    PEDIATRIC PULMONOLOGY, 2007, 42 (09) : 773 - 778
  • [10] Sweat test practice in pediatric pulmonology after introduction of cystic fibrosis newborn screening
    Céline Grimaldi
    François Brémont
    Michèle Berlioz-Baudoin
    Jacques Brouard
    Harriet Corvol
    Laure Couderc
    Guillaume Lezmi
    Isabelle Pin
    Isabelle Petit
    Philippe Reix
    Natacha Remus
    Cyril Schweitzer
    Caroline Thumerelle
    Jean-Christophe Dubus
    European Journal of Pediatrics, 2015, 174 : 1613 - 1620