Effect of table height displacement and patient center deviation on size-specific dose estimates calculated from computed tomography localizer radiographs

被引:0
作者
Kazuki Kuriyama
Kosuke Matsubara
Shu Hisahara
Yukie Nagata
Rumiko Nosaka
Rika Goto
Nobutaka Yanano
Kanichiro Shimizu
Tomokazu Shoji
机构
[1] The Jikei University Kashiwa Hospital,Department of Radiology
[2] Kanazawa University,Division of Health Sciences, Department of Quantum Medical Technology, Graduate Course of Medical Science and Technology, Graduate School of Medical Science
[3] Kanazawa University,Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences
[4] The Jikei University School of Medicine,Department of Radiology
[5] Kashiwa Hospital,Department of Radiology
[6] The Jikei University Hospital,undefined
来源
Physical and Engineering Sciences in Medicine | 2020年 / 43卷
关键词
Size-specific dose estimate; Miscentering; Table height; Computed tomography; Localizer radiograph;
D O I
暂无
中图分类号
学科分类号
摘要
The aim of this study is to evaluate the effect of table height displacement and patient center deviation along the z\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$z$$\end{document}-axis on size-specific dose estimate (SSDE) calculations based on computed tomography (CT) localizer radiographs in pediatric and adult abdominal CT examinations. CT localizer radiographs and CT axial images were acquired with table heights of − 5.0, − 2.5, 0.0 (center), 2.5, and 5.0 cm using two acrylic self-made phantoms filled with water. Water-equivalent diameters (Dw\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$${D}_{\mathrm{w}}$$\end{document}) were calculated from the CT localizer radiographs and CT axial images. Relative errors of SSDEs from the CT localizer radiographs to SSDEs from the CT axial images were calculated to evaluate the effect of table height displacement. Furthermore, patient center deviations and indices of SSDE overestimation were measured from the clinical data of 110 abdominal CT examinations. The relative errors of SSDEs in phantoms equivalent to 1-year-old and 20-year-old Japanese reference persons ranged from − 2.45% (table height of 50 mm) to + 1.88% (− 50 mm) and from − 4.22% (50 mm) to + 3.79% (− 50 mm), respectively. The largest center deviation in all patients ranged from − 43.1 to 21.5 mm (median: − 14.4 mm). The indices of SSDE overestimation for all patients ranged from − 16.2 to 6.9 mm (median: − 2.2 mm). We found that the effects of table height displacement and patient center deviation along the z\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$z$$\end{document}-axis on SSDEs calculated from CT localizer radiographs in pediatric phantoms were smaller compared to adult phantoms. In order to correct these patient center deviations, it is necessary to apply an appropriate correction technique in each section along the z\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$z$$\end{document}-axis.
引用
收藏
页码:665 / 672
页数:7
相关论文
共 44 条
  • [1] Brenner D(2001)Estimated risks of radiation-induced fatal cancer from pediatric CT AJR 176 289-296
  • [2] Elliston C(2011)CT dose index and patient dose: they are not the same thing Radiology 259 311-316
  • [3] Hall E(2012)Attenuation based estimation of patient size for the purpose of the size specific dose estimation in CT: part II. Implementation on abdomen and thorax phantoms using cross sectional CT images and scanned projection radiograph images Med Phys 39 6772-6778
  • [4] Berdon W(1981)A method for describing the doses delivered by transmission X-ray computed tomography Med Phys 8 488-495
  • [5] McCollough CH(2002)AAPM/RSNA physics tutorial for residents: topics in CT radiation dose in CT Radiographics 22 1541-1553
  • [6] Leng S(2015)Size-specific dose estimates for chest, abdominal, and pelvic CT: effect of intrapatient variability in water-equivalent diameter Radiology 276 184-190
  • [7] Yu L(2012)Comparison of topogram-based body size indices for CT dose consideration and scan protocol optimization Med Phys 39 3456-3465
  • [8] Cody DD(2019)Determination of appropriate conversion factors for calculating size-specific dose estimates based on X-ray CT scout images after miscentering correction Radiol Phys Technol 12 283-289
  • [9] Boone JM(2013)Assessment of patient dose from CT localizer radiographs Med Phys 40 084301-168
  • [10] McNitt-Gray MF(2018)A simple method for calibrating pixel values of the CT localizer radiograph for calculating water-equivalent diameter and size-specific dose estimate Radiat Prot Dosimetry 179 158-undefined