An ion pulse ionization chamber for spectrometric measurements of low surface α activities

被引:0
作者
Yu. M. Gavrilyuk
A. M. Gangapshev
V. V. Kazalov
V. V. Kuz’minov
S. I. Panasenko
S. S. Ratkevich
机构
[1] Russian Academy of Sciences,Baksan Neutrino Observatory, Institute for Nuclear Research
[2] Karazin National University,undefined
来源
Instruments and Experimental Techniques | 2009年 / 52卷
关键词
29.40.Cs; 91.67.Qr; 93.85.Np;
D O I
暂无
中图分类号
学科分类号
摘要
The design of a pulse ionization chamber for measuring the surface αactivity of various samples is described. When used together with the appropriate methods for recording and processing the signals, a low-background spectrometer based on this chamber is able to ensure a high sensitivity in direct measurements of ultralow surface α activities. This approach offers a solution to the problem of controlling superpure materials used in detectors for studying rare nuclear processes with a broken equilibrium in the decay chains of the uranium and thorium series. A procedure for modeling the shape of the current pulses according to the orientation of α-particle tracks and their location inside the chamber is described. Results obtained with an α source, as well as with copper and VM2000 light-reflecting film samples, are presented.
引用
收藏
页码:173 / 182
页数:9