Identification and quantification by 1H nuclear magnetic resonance spectroscopy of seven plasticizers in PVC medical devices
被引:0
|作者:
Stéphanie Genay
Frédéric Feutry
Morgane Masse
Christine Barthélémy
Valérie Sautou
Pascal Odou
Bertrand Décaudin
Nathalie Azaroual
机构:
[1] EA 7365 - GRITA - Groupe de Recherche sur les formes Injectables et les Technologies Associées,Univ. Lille, CHU Lille
[2] EA 4676,Clermont Université, Université d’Auvergne, CHU Clermont
[3] C-BIOSENSS,Ferrand
来源:
关键词:
Nuclear magnetic resonance;
Plasticizers;
Polyvinyl chloride;
Medical devices;
Biomaterials;
Polymers;
D O I:
暂无
中图分类号:
学科分类号:
摘要:
Medical devices are generally made of polyvinyl chloride plasticized by six authorized plasticizers as alternatives to di-(2-ethylhexyl)-phthalate (DEHP) classified as reprotoxic class 1b. These are acetyl tri-n-butyl citrate (ATBC), di-(2-ethylhexy) adipate (DEHA), di-(2-ethylhexyl) terephthalate (DEHT), di-isononyl cyclohexane-1,2-dicarboxylate (DINCH), di-isononyl phthalate (DINP), and tri-octyl trimellitate (TOTM). The main objective of this study was to propose a new method using 1H NMR spectroscopy to determine and quantify these seven plasticizers in PVC sheets, standard infusion tubings, and commercially available medical devices. Two techniques were compared: dissolution in deuterated tetrahydrofuran and extraction by deuterated chloroform. Plasticizer 1H NMR spectra were very similar in both deuterated solvents; dissolution and extraction provided similar results. The sensitivity of this method enabled us to detect and quantify the presence of minor plasticizers in PVC. In nine commercially available samples, the major plasticizer was identified and quantified by 1H NMR. In six samples, one, two, or three minor plasticizers were identified and also quantified. DEHP was detected in only one tubing. NMR is therefore very convenient for studying plasticizers contained in medical devices. Only small quantities of solvents and sample are required. It is not necessary to dilute samples to enter a quantification range, and it is sufficiently sensitive to detect contaminants.
引用
收藏
页码:1271 / 1280
页数:9
相关论文