A softening laminar electrode for recording single unit activity from the rat hippocampus

被引:0
作者
A. Zátonyi
G. Orbán
R. Modi
G. Márton
D. Meszéna
I. Ulbert
A. Pongrácz
M. Ecker
W. E. Voit
A. Joshi-Imre
Z. Fekete
机构
[1] Pázmány Péter Catholic University,Research Group for Implantable Microsystems, Faculty of Information Technology & Bionics
[2] University of Pannonia,Doctoral School of Chemical Engineering and Material Sciences
[3] Hungarian Academy of Sciences,Microsystems Laboratory, Institute for Technical Physics & Material Sciences, Centre for Energy Research
[4] Óbuda University Doctoral School on Materials Sciences & Technologies,Institute of Cognitive Neuroscience and Psychology, Research Centre for Natural Sciences
[5] Hungarian Academy of Sciences,Advanced Polymer Research Laboratory
[6] The University of Texas at Dallas,undefined
来源
Scientific Reports | / 9卷
关键词
D O I
暂无
中图分类号
学科分类号
摘要
Softening neural implants that change their elastic modulus under physiological conditions are promising candidates to mitigate neuroinflammatory response due to the reduced mechanical mismatch between the artificial interface and the brain tissue. Intracortical neural probes have been used to demonstrate the viability of this material engineering approach. In our paper, we present a robust technology of softening neural microelectrode and demonstrate its recording performance in the hippocampus of rat subjects. The 5 mm long, single shank, multi-channel probes are composed of a custom thiol-ene/acrylate thermoset polymer substrate, and were micromachined by standard MEMS processes. A special packaging technique is also developed, which guarantees the stable functionality and longevity of the device, which were tested under in vitro conditions prior to animal studies. The 60 micron thick device was successfully implanted to 4.5 mm deep in the hippocampus without the aid of any insertion shuttle. Spike amplitudes of 84 µV peak-to-peak and signal-to-noise ratio of 6.24 were achieved in acute experiments. Our study demonstrates that softening neural probes may be used to investigate deep layers of the rat brain.
引用
收藏
相关论文
共 86 条
[1]  
Fekete Z(2015)Recent advances in silicon-based neural microelectrodes and microsystems: a review Sensors and Actuators B: Chemical 215 300-315
[2]  
Jorfi M(2014)Progress towards biocompatible intracortical microelectrodes for neural interfacing applications Journal of Neural Engineering 12 011001-169
[3]  
Skousen JL(2016)Dexamethasone retrodialysis attenuates microglial response to implanted probes Biomaterials 87 157-9172
[4]  
Weder C(2016)Neurobiochemical changes in the vicinity of a nanostructured neural implant Scientific Reports 6 9163-3259
[5]  
Capadona JR(2010)A comparison of the tissue response to chronically implanted Parylene-C-coated and uncoated planar silicon microelectrode arrays in rat cortex Biomaterials 31 70-1223
[6]  
Kozai TD(2015)Experimental study on the mechanical interaction between silicon neural microprobes and rat dura mater during insertion Journal of Materials Science: Materials in Medicine 26 3250-1374
[7]  
Jaquins-Gerstl AS(2011)Effect of insertion speed on tissue response and insertion mechanics of a chronically implanted silicon-based neural probe IEEE Transactions on Biomedical Engineering 58 1214-3479
[8]  
Vazquez AL(2017)Multifunctional soft implants to monitor and control neural activity in the central and peripheral nervous system: A review Sensors and Actuators B: Chemical 243 036002-1202
[9]  
Michael AC(2015)Compliant intracortical implants reduce strains and strain rates in brain tissue Journal of Neural Engineering 12 023009-11
[10]  