Effects of the Design of a Rotary Blood Pump on Hemocompatibility

被引:0
作者
M. V. Denisov
M. Walter
S. Leonhard
D. V. Telyshev
机构
[1] Institute of Biomedical Systems,
[2] National Research University of Electronic Technology (MIET),undefined
[3] Rheinisch-Westfälische Technische Hochschule Aachen,undefined
来源
Biomedical Engineering | 2021年 / 54卷
关键词
D O I
暂无
中图分类号
学科分类号
摘要
The paper shows how hemocompatibility is affected by changes in the geometry of the flow part of a rotary blood pump (RBP) for supporting patients in the terminal stage of heart failure. Numerical simulation made it possible to obtain a family of head–flow curves for six geometrical RBP models differing in terms of the geometry of the impeller blade and the distance from the impeller to the diffusor. The effects of changes in geometry on the volume in which shear stresses are elevated were evaluated at the working point of 4.5 L/min at 80 mm Hg. The values for the minimum, maximum, and average residence times of blood particles in the pump domain were obtained. Numerical simulation of the flow was run in the Fluent ANSYS 19.0 software.
引用
收藏
页码:327 / 332
页数:5
相关论文
共 29 条
[1]  
Miller LW(2013)Cost of ventricular assist devices: Can we afford the progress? Circulation 127 743-748
[2]  
Guglin M(2013)Orthotopic heart transplant versus left ventricular assist device: A national comparison of cost and survival J. Thorac. Cardiovasc. Surg. 145 566-574
[3]  
Rogers J(2015)Development of left ventricular assist devices as the most effective acute heart failure therapy Biomed. Eng. 48 328-330
[4]  
Mulloy DP(2016)Optimisation of the Sputnik-VAD design Int. J. Artif. Org. 39 407-414
[5]  
Bhamidipati CM(2017)The effect of rotor geometry on the H–Q curves of the Sputnik implantable pediatric rotary blood pump Biomed. Eng. 50 420-424
[6]  
Stone ML(2017)Development of medical and technical requirements and simulation of the flow-pressure characteristics of the Sputnik pediatric rotary blood pump Biomed. Eng. 50 296-299
[7]  
Ailawadi G(2018)The progress in the novel pediatric rotary blood pump Sputnik development Artif. Org. 42 432-443
[8]  
Kron IL(1995)Model for a general mechanical blood damage prediction Artif. Org. 19 583-589
[9]  
Kern JA(2017)A cellular model of shear-induced hemolysis Artif. Org. 41 1-12
[10]  
Petukhov DS(undefined)undefined undefined undefined undefined-undefined