Evaluation of different near-infrared spectroscopy technologies for assessment of tissue oxygen saturation during a vascular occlusion test

被引:0
作者
Kevin Steenhaut
Koen Lapage
Thierry Bové
Stefan De Hert
Annelies Moerman
机构
[1] Ghent University Hospital,Department of Anesthesiology
[2] Ghent University Hospital,Department of Cardiac Surgery
来源
Journal of Clinical Monitoring and Computing | 2017年 / 31卷
关键词
Near-infrared spectroscopy; Tissue oxygen saturation; Vascular occlusion test;
D O I
暂无
中图分类号
学科分类号
摘要
An increasing number of NIRS devices are used to provide measurements of peripheral tissue oxygen saturation (StO2). The aim of the present study is to test the hypothesis that despite technological differences between devices, similar trend values will be obtained during a vascular occlusion test. The devices compared are NIRO-200NX, which measures StO2 and oxyhemoglobin by spatially resolved spectroscopy and the Beer–Lambert law, respectively, and INVOS 5100C and Foresight Elite, which both measure StO2 with the Beer–Lambert law, enhanced with the spatial resolution technique. Forty consenting adults scheduled for CABG surgery were recruited. The respective sensors of the three NIRS devices were applied over the brachioradial muscle. Before induction of anesthesia, 3 min of ischemia were induced by inflating a blood pressure cuff at the upper arm, whereafter cuff pressure was rapidly released. Tissue oxygenation measurements included baseline, minimum and maximum values, desaturation and resaturation slopes, and rise time. Comparisons between devices were performed with the Kruskal–Wallis test with post hoc Mann–Whitney pairwise comparisons. Agreement was evaluated using Bland–Altman plots. Oxyhemoglobin measured with NIRO responded faster than the other NIRS technologies to changes in peripheral tissue oxygenation (20 vs. 27–40 s, p ≤ 0.01). When comparing INVOS with Foresight, oxygenation changes were prompter (upslope 311 [92–523]%/min vs. 114[65–199]%/min, p ≤ 0.01) and more pronounced (minimum value 36 [21–48] vs. 45 [40–51]%, p ≤ 0.01) with INVOS. Significant differences in tissue oxygen saturation measurements were observed, both within the same device as between different devices using the same measurement technology.
引用
收藏
页码:1151 / 1158
页数:7
相关论文
共 72 条
  • [1] Scheeren TW(2016)Journal of Clinical Monitoring and Computing 2015 end of year summary: tissue oxygenation and microcirculation J Clin Monit Comput 30 141-146
  • [2] Moerman A(2010)Near-infrared spectroscopy (NIRS) monitoring in contemporary anesthesia and critical care Acta Anaesth Belg 61 185-194
  • [3] Wouters P(2014)Peripheral tissue oximetry: comparing three commercial near-infrared spectroscopy oximeters on the forearm J Clin Monit Comput 28 149-155
  • [4] Hyttel-Sorensen S(2011)Tissue oximetry: a comparison of mean values of regional tissue saturation, reproducibility and dynamic range of four NIRS-instruments on the human forearm Biomed Opt Express 2 3047-3057
  • [5] Hessel TW(2009)Assessment of tissue oxygen saturation during a vascular occlusion test using near-infrared spectroscopy: the role of probe spacing and measurement site studied in healthy volunteers Crit Care 13 S4-10
  • [6] Greisen G(2013)Influence of cutaneous and muscular circulation on spatially resolved versus standard Beer–Lambert near-infrared spectroscopy Physiol Rep 1 1-1655
  • [7] Hyttel-Sorensen S(2011)Inconsistent detection of changes in cerebral blood volume by near infrared spectroscopy in standard clinical tests J Appl Physiol 110 1646-438
  • [8] Sorensen LC(2010)Depth sensitivity analysis of functional near-infrared spectroscopy measurement using three-dimensional Monte Carlo modelling-based magnetic resonance imaging Lasers Med Sci 25 431-175
  • [9] Riera J(2013)The effect of basic assumptions on the tissue oxygen saturation value of near infrared spectroscopy Adv Exp Med Biol 765 169-265
  • [10] Greisen G(2013)Relation between mixed venous oxygen saturation and cerebral oxygen saturation measured by absolute and relative near-infrared spectroscopy during off-pump coronary artery bypass grafting Br J Anaesth 110 258-278