Assessment of dosimetric leaf gap correction factor in Mobius3D commissioning affected by couch top

被引:0
|
作者
Min Cheol Han
Jihun Kim
Chae-Seon Hong
Kyung Hwan Chang
Su Chul Han
Kwangwoo Park
Dong Wook Kim
Myoung Kyu Park
Yu Yun Noh
Jin Sung Kim
机构
[1] Yonsei University College of Medicine,Department of Radiation Oncology
[2] Yonsei Cancer Center,Department of Radiation Oncology
来源
Physical and Engineering Sciences in Medicine | 2020年 / 43卷
关键词
Mobius; Couch effect; Dosimetric leaf gap; Mobius verification phantom;
D O I
暂无
中图分类号
学科分类号
摘要
This study assesses the dosimetric leaf gap (DLG) correction factor in Mobius3D commissioning affected by a couch top platform and calculates the optimal DLG value according to the point dose difference function. DLG optimizations were performed for 3 LINAC machines and a total of 30 patient volumetric modulated arc therapy plans (i.e., 10 plans per each LINAC). Point dose calculations were performed using an automatic dose calculation system in Mobius3D as well as Mobis3D calculation using a Mobius Verification Phantom (MVP)-based quality assurance plan with a carbon fiber couch top. Subsequently, the results were compared with measurement data. The averaged point dose measured for the MVP with a couch top decreased by approximately 2% relative to that without the couch top. The average of the optimal DLG factors increased by 1.153 mm due to the couch top effect for a dose decrease of 2% at the measured point. In the procedure of Mobius beam commissioning, users should adjust the DLG correction factor using a specific phantom (including MVP) with a couch top structure. If the DLG optimization were performed by using MVP automatic dose calculation system, the factor should be increased by approximately 1.2 mm per 2% dose difference considering user’s couch top effect.
引用
收藏
页码:1069 / 1075
页数:6
相关论文
共 15 条
  • [1] Assessment of dosimetric leaf gap correction factor in Mobius3D commissioning affected by couch top
    Han, Min Cheol
    Kim, Jihun
    Hong, Chae-Seon
    Chang, Kyung Hwan
    Han, Su Chul
    Park, Kwangwoo
    Kim, Dong Wook
    Park, Myoung Kyu
    Noh, Yu Yun
    Kim, Jin Sung
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (03) : 1069 - 1075
  • [2] Characterization of the Limitations of the Mobius3D Dosimetric Leaf Gap Offset Correction
    Shepard, A.
    Hansen, J.
    Belanger, M.
    Frigo, S.
    MEDICAL PHYSICS, 2020, 47 (06) : E716 - E717
  • [3] The Commissioning of DLG Correction Factor for Mobius3D
    Wang, B.
    Yang, B.
    Qiu, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E663 - E663
  • [4] Commissioning of Mobius3D for Linac-Based SRS and Determination of the Dosimetric-Leaf-Gap
    McCarroll, R.
    Sabouri, P.
    Lamichhane, N.
    Poirier, Y.
    Yi, B.
    Chen, S.
    Guerrero, M.
    MEDICAL PHYSICS, 2019, 46 (06) : E612 - E612
  • [5] Investigating the Dosimetric Leaf Gap Correction Factor of Mobius3D Dose Calculation for Volumetric-modulated Arc Radiotherapy Plans
    Sawapabmongkon, Thitipong
    Changkaew, Pimolpun
    Puttanawarut, Chanon
    Tangboonduangjit, Puangpen
    Khachonkham, Suphalak
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (02) : 261 - 269
  • [6] Commissioning of the Mobius3D independent dose verification system for TomoTherapy
    Kodama, Takumi
    Saito, Yoshihiro
    Hatanaka, Shogo
    Hariu, Masatsugu
    Shimbo, Munefumi
    Takahashi, Takeo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (05): : 12 - 20
  • [7] Commissioning and clinical implementation of Mobius3D and MobiusFX: Experience on multiple linear accelerators
    Kim, Jihun
    Han, Min Cheol
    Park, Kwangwoo
    Chang, Kyung Hwan
    Kim, Dong Wook
    Hong, Chae-Seon
    Kim, Jin Sung
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 80 : 1 - 9
  • [8] Estimation of Dosimetric Leaf Gap Correction Factor as Part of Commissioning a Log File Based QA System
    Sreekala, U.
    Shine, N. S.
    Raghukumar, P.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [9] Assessment of the Dosimetric Performance of the Mobius3D against Portal Dose Measurements in Patient-specific Quality Assurance
    Labagnoy, Yasmine Joy M.
    Oonsiri, Sornjarod
    Vimolnoch, Mananchaya
    Kingkaew, Sakda
    JOURNAL OF MEDICAL PHYSICS, 2023, 48 (04) : 350 - 357
  • [10] Assessment of log-based fingerprinting system of Mobius3D with Elekta linear accelerators
    Noh, Yu-Yun
    Kim, Jihun
    Kim, Jin Sung
    Shin, Han-Back
    Han, Min Cheol
    Suh, Tae Suk
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (02):