The reduction of artifacts due to metal hip implants in CT-attenuation corrected PET images from hybrid PET/CT scanners

被引:0
作者
John A. Kennedy
Ora Israel
Alex Frenkel
Rachel Bar-Shalom
Haim Azhari
机构
[1] Technion Israel Institute of Technology,Faculty of Biomedical Engineering
[2] Rambam Health Care Campus,Department of Nuclear Medicine
[3] Technion Israel Institute of Technology,Bruce Rappaport Faculty of Medicine
来源
Medical & Biological Engineering & Computing | 2007年 / 45卷
关键词
PET/CT; Artifact; Hip prosthesis; Attenuation correction;
D O I
暂无
中图分类号
学科分类号
摘要
CT beam hardening artifacts near metal hip implants may erroneously enhance or diminish radiotracer uptake following CT attenuation correction (AC) of PET images. An artifact reduction algorithm (ARA) was developed to reduce metal artifacts in CT-based AC-PET. The algorithm employed a Bayes classifier to identify beam-hardening artifacts, followed by a partial correction of the attenuation map. ARA was implemented on phantom and patient 18F-FDG studies using a clinical PET/CT scanner. In phantom studies ARA successfully removed two artifacts of erroneously elevated uptake near a stainless steel hip prosthesis which were depicted in the standard CT-AC PET. ARA has also identified two targets absent on the scanner PET images. Target-to-background ratios were 1.5–3 times higher for ARA-PET than scanner images. In a patient study, metal artifacts were of lower intensity in ARA-PET as compared to standard images. Potentially, ARA may improve detectability of small lesions located near metal hip implants.
引用
收藏
页码:553 / 562
页数:9
相关论文
共 113 条
[11]  
Yefremov N(2002)Is quantitation necessary for oncological PET studies? Against Eur J Nucl Med Mol Imaging 29 135-138
[12]  
Guralnik L(1994)Accelerated image reconstruction using ordered subsets of projection data IEEE Trans Med Imag 13 601-609
[13]  
Gaitini D(1987)Reduction of CT artifacts caused by metallic implants Radiology 164 576-577
[14]  
Frenkel A(2003)Impact of metallic dental implants on CT-based attenuation correction in a combined PET/CT scanner Eur Radiol 13 724-728
[15]  
Kuten A(2002)CT vs Eur J of Nucl Med 29 246-350
[16]  
Beyer T.(2005)Ge attenuation correction in a combined PET/CT system: evaluation of the effect of lowering the CT tube current J Nucl Med 46 444-449
[17]  
Townsend D.W.(1998)The diabetic foot: initial experience with Med Phys 25 2046-2053
[18]  
Brun T(2003)F-FDG PET/CT IEEE Nucl Sci Symp Conf Rec 3 2220-2221
[19]  
Kinahan PE(2006)Attenuation correction for a combined PET/CT scanner Am J Roentgenol 186 1458-1467
[20]  
Charron M(2005)Evaluation of method to minimize the effect of X-ray contrast in PET-CT attenuation correction Conf Rec of the 2005 IEEE Nuc Sci Symp 4 1895-1899