The measuring converter is based on the principle of tracking the balancing of the separated informative value of the interval probed by an acoustic pulse in the studied medium at a given basic distance by a pulse period divisible by the period of pulses of a frequency-controlled generator. A high accuracy is reached by excluding an inconclusive (spurious) time delay of the acoustic and electric signals in the acoustic converter and electronic operational units. The acoustic base-line distance of the measuring converter is 0.1 m, the measurement range is 1400–1600 m/s, the output signal has a frequency of 140–160 kHz, the absolute basic error is ≤ ±0.015 m/s, and the time to reach a steady-state output signal is ≤0.6 s.