A quartz crystal microbalance sensor for the detection of formic acid vapors

被引:0
作者
C. García-Verdugo-Caso
J. L. Hidalgo-Hidalgo-de-Cisneros
I. Naranjo-Rodríguez
D. Bellido-Milla
机构
[1] Department of Analytical Chemistry,
[2] Faculty of Sciences,undefined
[3] University of Cádiz. Apdo. 40,undefined
[4] 11510 Puerto Real,undefined
[5] Cádiz,undefined
[6] Spain e-mail: jluis.hidalgo@uca.es,undefined
来源
Fresenius' Journal of Analytical Chemistry | 2001年 / 370卷
关键词
Molecular Mass; Formic Acid; Carboxylic Acid; Linear Range; Mechanical Response;
D O I
暂无
中图分类号
学科分类号
摘要
A new quartz crystal microbalance sensor is developed to determine formic acid at low concentrations. Four previously selected polymers with acid–base characteristics were tested as possible coatings. Polyoxyethylene bis [amine] presented the best results. The sensor is rapid, sensitive [0.67 Hz/(mg/m3)], and reversible at low concentrations. The detection limit for formic acid (7.2 mg/m3) is comparable with the short term exposure limit and the threshold limit values. It presents a fast mechanical response to pressure changes, so that it can be quickly used in different environments and situations. The sensor also shows a good stability in a temperature range typical of work atmospheres (16–36 °C). It has a wide linear range (7.2–911.2 mg/m3) and a long useful time. It is also applicable to other low molecular mass carboxylic acids such as acetic acid.