Proximal outflow graft extrinsic compression of the Heartmate 3 centrifugal flow left ventricular assist device resulting in pump dysfunction

被引:0
作者
Julien Guihaire
Martin Kloeckner
Marc Antoine Isorni
机构
[1] University of Paris Sud,Cardiac Surgery Department, Marie Lannelongue Hospital
[2] University of Paris Sud,Research and Innovation Unit, Marie Lannelongue Hospital
[3] University of Paris Sud,Inserm U999, PAH: Pathophysiology and Novel Therapies
[4] University of Paris Sud,Department of Diagnostic and Therapeutic Imaging, Marie Lannelongue Hospital
来源
Journal of Artificial Organs | 2020年 / 23卷
关键词
Left ventricular assist device; Complication; Thrombosis;
D O I
暂无
中图分类号
学科分类号
摘要
We herein report an unusual life-threatening complication with the Hearmate 3 centrifugal ventricular assist device: proximal outflow graft extrinsic compression due to external thrombus leading to pump dysfunction and urgent heart transplantation. Low flow alarms without other abnormalities lead to the diagnosis of outflow occlusion. There was no evidence for a twist of the outflow graft, and no evidence for outflow thrombosis. The location of the outflow occlusion was particularly difficult to diagnose due to metallic artifacts around the pump. This report underscores that this complication should be screened in case of persistent low-flow alarms with the Heartmate 3 centrifugal pump.
引用
收藏
页码:180 / 182
页数:2
相关论文
共 22 条
[1]  
Mehra MR(2017)A fully magnetically levitated circulatory pump for advanced heart failure N Engl J Med 376 440-450
[2]  
Naka Y(2018)Two-year outcomes with a magnetically levitated cardiac pumpin heart failure N Engl J Med 378 1386-1395
[3]  
Uriel N(2017)Outflow graft occlusion of the heartmate left ventricular assist device Circ Heart Fail 10 e004275-474
[4]  
Mehra MR(2019)Surgical treatment of late left ventricular assist device outflow obstruction due to extrinsic compression J Heart Lung Transplant 38 472-1284
[5]  
Goldstein DJ(2018)A tale of the twist in the outflow graft: an analysis from the MOMENTUM 3 trial J Heart Lung Transpl 37 1281-475
[6]  
Uriel N(2019)Avoidance of outflow graft extrinsic compression in polytetrafluoroethylene reinforcements of the bend relief component J Heart Lung Transplant 38 474-undefined
[7]  
Duero Posada JG(undefined)undefined undefined undefined undefined-undefined
[8]  
Moayedi Y(undefined)undefined undefined undefined undefined-undefined
[9]  
Alhussein M(undefined)undefined undefined undefined undefined-undefined
[10]  
Rajagopal K(undefined)undefined undefined undefined undefined-undefined