Comparison of two generalized transfer functions for measuring central systolic blood pressure by an oscillometric blood pressure monitor

被引:0
作者
Y-T Shih
H-M Cheng
S-H Sung
W-C Hu
C-H Chen
机构
[1] Chung Yuan Christian University,Department of Biomedical Engineering
[2] Taipei Veterans General Hospital,Department of Medical Research and Education
[3] Taipei Veterans General Hospital,Department of Medicine
[4] Cardiovascular Research Center,Department of Medicine
[5] National Yang-Ming University,Department of Public Health
[6] National Yang-Ming University,undefined
[7] National Yang-Ming University,undefined
来源
Journal of Human Hypertension | 2013年 / 27卷
关键词
generalized transfer function; pulse volume plethysmography; oscillometry; central blood pressure;
D O I
暂无
中图分类号
学科分类号
摘要
Central aortic systolic blood pressure (SBP-C) can be estimated from a cuff oscillometric waveform derived during the pulse volume plethysmography (PVP) by applying a device-specific aortic pressure-to-PVP waveform-generalized transfer function (A2PGTF). The present study compared the performance of an aortic-to-brachial pressure waveforms generalized transfer function (A2BGTF), which is independent of any PVP devices, with an A2PGTF. Generalized transfer function of aortic-to-brachial (A2BGTF) and aortic-to-PVP (A2PGTF) were generated from the simultaneously obtained central aortic and brachial pressure waveforms recorded by a high-fidelity dual pressure sensor catheter, and the PVP waveform recorded by a customized noninvasive blood pressure monitor during cardiac catheterization in 40 patients, and were then applied in another 100 patients with simultaneously recorded invasive aortic pressure and noninvasively calibrated (using cuff SBP and diastolic blood pressures) PVP waveforms. The mean difference±s.d. between the noninvasively estimated and invasively recorded SBP-C was −2.1±7.7 mm Hg for A2BGTF, which was not greater than that of −3.0±7.7 mm Hg for A2PGTF (P<0.01). In conclusion, SBP-C can be measured reliably using a noninvasive blood pressure monitor by applying either an A2PGTF or A2BGTF to a noninvasively calibrated PVP waveform. The performance of an A2BGTF is not inferior to that of an A2PGTF.
引用
收藏
页码:204 / 210
页数:6
相关论文
共 129 条
[1]  
Segers P(2009)Amplification of the pressure pulse in the upper limb in healthy, middle-aged men and women Hypertension 54 414-420
[2]  
Mahieu D(2010)Cuff and aortic pressure differences during dobutamine infusion: a study of the effects of systolic blood pressure amplification Am Heart J 159 399-405
[3]  
Kips J(2003)Spurious systolic hypertension of youth: fit young men with elastic arteries Am J Hypertens 16 229-232
[4]  
Rietzschel E(2010)Estimation of central systolic blood pressure using an oscillometric blood pressure monitor Hypertens Res 33 592-599
[5]  
De Buyzere M(2006)Validation of a generalized transfer function to noninvasively derive central blood pressure during exercise Hypertension 47 1203-1208
[6]  
De Bacquer D(2010)Prediction of cardiovascular events and all-cause mortality with central haemodynamics: a systematic review and meta-analysis Eur Heart J 31 1865-1871
[7]  
Smulyan H(2010)Central haemodynamics and clinical outcomes: going beyond brachial blood pressure? Eur Heart J 31 1819-1822
[8]  
Mukherjee R(2009)Non-invasive methods and techniques for central blood pressure estimation: procedures, validation, reproducibility and limitations Curr Pharm Des 15 245-253
[9]  
Sheehe PR(2011)Blood pressure measurement: retrospective and prospective views Am J Hypertens. 24 628-634
[10]  
Safar ME(2007)Central pressure more strongly relates to vascular disease and outcome than does brachial pressure: the Strong Heart Study Hypertension 50 197-203