Mechanical strength assessment of a drilled hole in the contralateral cortex at the end of the open wedge for high tibial osteotomy

被引:9
作者
Diffo Kaze A. [1 ,2 ,3 ,4 ]
Maas S. [1 ,4 ]
Hoffmann A. [2 ,3 ,4 ]
Pape D. [2 ,3 ,4 ]
机构
[1] Faculty of Science, Technology and Communication, University of Luxembourg, 6, rue R. Coudenhove-Kalergi, Luxembourg
[2] Department of Orthopedic Surgery, Centre Hospitalier de Luxembourg, Luxembourg
[3] Sports Medicine Research Laboratory, Public Research Centre for Health, Luxembourg, Centre Médical de la Fondation Norbert Metz, 76 rue d’Eich, Luxembourg
[4] Cartilage Net of the Greater Region, Homburg/Saar
关键词
Biomechanics; Correction angle; Drill hole; Elongation at break; Finite element analysis; High tibial osteotomy (HTO); Opposite cortical fracture; Osteoarthritis (OA); Static strength;
D O I
10.1186/s40634-017-0098-0
中图分类号
学科分类号
摘要
Background: This study aimed to investigate, by means of finite element analysis, the effect of a drill hole at the end of a horizontal osteotomy to reduce the risk of lateral cortex fracture while performing an opening wedge high tibial osteotomy (OWHTO). The question was whether drilling a hole relieves stress and increases the maximum correction angle without fracture of the lateral cortex depending on the ductility of the cortical bone. Methods: Two different types of osteotomy cuts were considered; one with a drill hole (diameter 5 mm) and the other without the hole. The drill holes were located about 20 mm distally to the tibial plateau and 6 mm medially to the lateral cortex, such that the minimal thickness of the contralateral cortical bone was 5 mm. Based on finite element calculations, two approaches were used to compare the two types of osteotomy cuts considered: (1) Assessing the static strength using local stresses following the idea of the FKM-guideline, subsequently referred to as the “FKM approach” and (2) limiting the total strain during the opening of the osteotomy wedge, subsequently referred to as “strain approach”. A critical opening angle leading to crack initiation in the opposite lateral cortex was determined for each approach and was defined as comparative parameter. The relation to bone aging was investigated by considering the material parameters of cortical bones from young and old subjects. Results: The maximum equivalent (von-Mises) stress was smaller for the cases with a drill hole at the end of the osteotomy cut. The critical angle was approximately 1.5 times higher for the specimens with a drill hole compared to those without. This corresponds to an average increase of 50%. The calculated critical angle for all approaches is below 5°. The critical angle depends on the used approach, on patient’s age and assumed ductility of the cortical bone. Conclusions: Drilling a hole at the end of the osteotomy reduces the stresses in the lateral cortex and increases the critical opening angle prior to cracking of the opposite cortex in specimen with small correction angles. But the difference from having a drill hole or not is not so significant, especially for older patients. The ductility of the cortical bone is the decisive parameter for the critical opening angle. © 2017, The Author(s).
引用
收藏
相关论文
共 20 条
[1]  
Agneskirchner J.D., Freiling D., Hurschler C., Lobenhoffer P., Primary stability of four different implants for opening wedge high tibial osteotomy, Knee Surg Sports Traumatol Arthrosc, 14, pp. 291-300, (2006)
[2]  
Beillas P., Begeman P.C., Yang K.H., King A.I., Arnoux P.J., Kang H.S., Kayvantash K., Brunet C., Prasad P., Lower Limb: Advanced FE Model and New Experimental Data, Stapp Car Crash J, 45, pp. 469-494, (2001)
[3]  
Evans F.G., Mechanical Properties and Histology of Cortical Bone from Younger and Older Men, Anat Rec, 185, pp. 1-12, (1976)
[4]  
Jo H.S., Park J.S., Byun J.H., Lee Y.B., Choi Y.L., The effects of different hinge positions on posterior tibial slope in medial open-wedge high tibial osteotomy, Knee Surg Sports Traumatol Arthrosc, pp. 1-8, (2017)
[5]  
Kessler O.C., Jacob H.A., Romero J., Avoidance of medial cortical fracture in high tibial osteotomy: improved technique, Clin Orthop Relat Res, 395, pp. 180-185, (2002)
[6]  
Online Materials Information Ressource.
[7]  
McCalden R.W., McGeough J.A., Barker M.B., Court-Brown C.M., Age-related changes in the tensile properties of cortical bone. The relative importance of changes in porosity, mineralization, and microstructure, J Bone Joint Surg Am, 75, pp. 1193-1205, (1993)
[8]  
Ott I., Ein Materialgesetz zur Beschreibung des Alterungsprozesses kortikaler Knochen, Dissertation, (2007)
[9]  
Ott I., Kienzler R., Schroder R., Aging in the cortical bone: a constitutive law and its application, Arch Appl Mech, 80, pp. 527-541, (2010)
[10]  
Pape D., Adam F., Rupp S., Seil R., Kohn D., Stabilität, Knochenheilung und Korrrekturverlust nach valgisierender Tibiakopfumstellung: Eine Röntgenstereometrieanalyse, Orthopade, 33, pp. 208-217, (2004)