Wear on joint prostheses is the leading cause of premature failure leading to revision surgery. Tests conducted on mechanical wear simulator show poor performance of prosthetic hip components manufactured using high-quality raw materials. Aiming to understand the phenomena that limit the longevity of joint prostheses, the manufacturing processes of two separate companies were investigated. Results obtained suggest that the lack of control during the manufacturing of the parts may be the reason for the low performance shown by the prostheses tested in the wear simulator. Out-of-roundness and roughness of finished surfaces is pointed out as the main manufacturing failure in production lines tested. Metal femoral heads showed deviations from roundness of up to 32.41 μm (σ = 8.869) and Ra = 0.295 ± 0.03, while acetabular cups in ultra-high molecular weight polyethylene showed deviations of up to 166.31 μm (σ = 53.982) and Ra = 0.031 ± 0.02. The out-of-roundness contributes to reducing the contact area between rolling surfaces, increasing tension in regions where premature wear on parts occurs with frequency. This study points out the common faults found in manufacturing processes and suggest changes to increase the longevity of total hip arthroplasty.
引用
收藏
页码:787 / 795
页数:8
相关论文
共 106 条
[1]  
Hughes PE(2002)Hip anatomy and biomechanics in the athlete Sports Med Arthrosc Rev 10 103-114
[2]  
Hsu JC(1971)Patterns of cartilage stiffness on normal and degenerate human femoral heads J Biomech 4 597-10
[3]  
Matava MJ(1976)Force actions transmitted by joints in the human body Proc Royal Soci Lon 192 163-undefined
[4]  
Kempson GE(2001)Hip contact forces and gait patterns from routine activities J Biomech 34 859-undefined
[5]  
Spivey CJ(1999)Current concepts review—wear in total hip and knee replacements J Bone Joint Surg 81 115-undefined
[6]  
Swanson SAV(1976)Principles and practice in hip replacement Royal Soc Lond 192 191-undefined
[7]  
Freeman MAR(1968)Total hip replacement Proc Royal Soc Lond 61 665-undefined
[8]  
Paul JP(2005)In vitro and in vivo biological responses to a novel radiopacifying agent for bone cement J Royal Soc Interface 2 71-undefined
[9]  
Bergmann G(2003)Bearing surface options for total hip replacement in young patients J Bone Joint Surg 85 1366-undefined
[10]  
Deuretzbacher G(1992)Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space J Bone Joint Surg 74 849-undefined