New computer-guided scanner for improving CO2 laser-assisted microincision

被引:0
作者
Marc Remacle
Faridah Hassan
David Cohen
Georges Lawson
Monique Delos
机构
[1] Louvain University Hospital at Mont-Godinne,Department of Otorhinolaryngology Head and Neck Surgery
[2] Louvain University Hospital at Mont-Godinne,Department of Anatomic Pathology
[3] Shaare Zedek Medical Center,Department of Otorhinolaryngology Head and Neck Surgery
来源
European Archives of Oto-Rhino-Laryngology and Head & Neck | 2005年 / 262卷
关键词
Computer-guided scanner; CO; laser-assisted microsurgery; Laser-produced coagulation;
D O I
暂无
中图分类号
学科分类号
摘要
The objective was to clinically test a new computer-guided scanner designed for CO2 laser-assisted microincision. The scanner-assisted beam travels across the target as a straight or curved incision line. Line length and beam penetration can be adjusted. The studied population, 155 cases, encompassed benign lesions as well as early cancers of the larynx. Operating time was compared with that required for similar operations performed with the Acuspot micromanipulator. Laser-produced coagulation thickness at the incision was measured on 41 operative specimens. The scanner-assisted incision and dissection were more accurate and required up to 30% less time than with a manually guided beam. Postoperative follow-up was straightforward. The coagulation thickness was less than 10 µm for phonomicrosurgery and less than 20 µm for other surgical procedures. The scanner-assisted incision is more accurate than that attained manually.
引用
收藏
页码:113 / 119
页数:6
相关论文
empty
未找到相关数据