Wireless non-invasive continuous respiratory monitoring with FMCW radar: a clinical validation study

被引:0
作者
K. van Loon
M. J. M. Breteler
L. van Wolfwinkel
A. T. Rheineck Leyssius
S. Kossen
C. J. Kalkman
B. van Zaane
L. M. Peelen
机构
[1] University Medical Center Utrecht,Department of Anesthesiology
[2] ZGT Hospital,Department of Intensive Care
[3] Radar Technology at TNO,Julius Center for Health Sciences and Primary Care
[4] University Medical Center Utrecht,undefined
来源
Journal of Clinical Monitoring and Computing | 2016年 / 30卷
关键词
Radar; Respiratory rate; Monitoring; Remote;
D O I
暂无
中图分类号
学科分类号
摘要
Altered respiratory rate is one of the first symptoms of medical conditions that require timely intervention, e.g., sepsis or opioid-induced respiratory depression. To facilitate continuous respiratory rate monitoring on general hospital wards a contactless, non-invasive, prototype monitor was developed using frequency modulated continuous wave radar. We aimed to study whether radar can reliably measure respiratory rate in postoperative patients. In a diagnostic cross-sectional study patients were monitored with the radar and the reference monitor (pneumotachograph during mechanical ventilation and capnography during spontaneous breathing). Eight patients were included; yielding 796 min of observation time during mechanical ventilation and 521 min during spontaneous breathing. After elimination of movement artifacts the bias and 95 % limits of agreement for mechanical ventilation and spontaneous breathing were −0.12 (−1.76 to 1.51) and −0.59 (−5.82 to 4.63) breaths per minute respectively. The radar was able to accurately measure respiratory rate in mechanically ventilated patients, but the accuracy decreased during spontaneous breathing.
引用
收藏
页码:797 / 805
页数:8
相关论文
共 74 条
[11]  
Sumner A(1965)An algorithm for the machine calculation of complex fourier series Math Comput 19 297-301
[12]  
Buist M(2007)ViEWS–towards a national early warning score for detecting adult inpatient deterioration Resuscitation 81 932-937
[13]  
Bernard S(2007)Agreement between methods of measurement with multiple observations per individual J Biopharm Stat 17 571-582
[14]  
Nguyen TV(1987)Evaluating clinical accuracy of systems for self-monitoring of blood glucose Diab Care 10 622-628
[15]  
Moore G(1999)Time course of changes in breathing pattern in morphine- and oxycodone-induced respiratory depression Anaesthesia 54 835-840
[16]  
Anderson J(2009)The effect of remifentanil on respiratory variability, evaluated with dynamic modeling J Appl Physiol 106 1038-1049
[17]  
Fieselmann JF(2010)Pattern recognition software and techniques for biological image analysis PLoS Comput Biol 6 e1000974-553
[18]  
Hendryx MS(2006)Robust overnight monitoring of human vital signs by a non-contact respiration and heartbeat detector IEEE Eng Med Biol Soc 49 537-undefined
[19]  
Helms CM(2009)Wavelet analysis of sensor signals for tool condition monitoring: a review and some new results Int J Mach Tools Manuf undefined undefined-undefined
[20]  
Wakefield DS(2013)Physical working principles of medical radar IEEE Trans Biomed Eng undefined undefined-undefined