Blood pressure differences between a mercury sphygmomanometer and two automatic devices

被引:0
|
作者
Lim H.K. [1 ,2 ]
Doh I. [1 ]
Ahn B. [1 ,2 ]
机构
[1] Korea Research Institute of Standards and Science, Seoul
[2] Korea University of Science and Technology, Seoul
关键词
Blood pressure; Cardiovascular disease; Measurement; Reference; Sphygmomanometer;
D O I
10.1007/s13534-015-0207-x
中图分类号
学科分类号
摘要
Purpose: Blood pressure (BP) is one of the most important pieces of information to diagnose cardiovascular diseases but the measurement readings could be easily changed by various factors. If these readings are wrong, it may ultimately lead to a misdiagnosis. In addition, there is a problem when there are different devices used to read blood pressure because these devices may give different readings. Methods: Four different BP measurement devices were used, including two mercury sphygmomanometers and two automatic digital devices. Thirty-eight healthy young people (mean age = 27.1±6.1 years) stayed one night at a clinical center and the BPs were measured from their right arms on the following day. Factors such as exercise, food, and smoking that may affect the BP were controlled and prohibited. All tests were repeated three times and all devices were calibrated before the test. Absolute BP value difference between a mercury sphygmomanometer and an automatic digital device was counted for the BP difference between devices ≥ 5 mmHg and ≥ 10 mmHg of each measurement. Results: Significant inter-device BP differences were found between a mercury sphygmomanometer and two automatic devices (p<0.01) while no significant difference was found between mercury sphygmomanometers (p>0.5); all subjects showed BP difference less than 5 mmHg between two mercury sphygmomanometers while only 32% and 38% of subjects showed BP difference less than 5 mmHg for two out of three repeated tests between a mercury sphygmomanometer and an automatic device, and even 26% ~ 36% of subjects showed BP difference ≥ 10 mmHg. Conclusions: This study result stresses that it is necessary to check the reliability of automatic blood pressure devices used in the hospitals. In addition, an easier and more convenient protocol should be developed for the routine calibration of automatic devices. © 2015, Korean Society of Medical and Biological Engineering and Springer.
引用
收藏
页码:296 / 301
页数:5
相关论文
共 50 条
  • [41] Environmental Mercury Exposure and Blood Pressure Among Nunavik Inuit Adults
    Valera, Beatriz
    Dewailly, Eric
    Poirier, Paul
    HYPERTENSION, 2009, 54 (05) : 981 - 986
  • [42] Chronic mercury exposure and blood pressure in children and adolescents: a systematic review
    Gallego-Vinas, Gema
    Ballester, Ferran
    Llop, Sabrina
    ENVIRONMENTAL SCIENCE AND POLLUTION RESEARCH, 2019, 26 (03) : 2238 - 2252
  • [43] Monitoring with wearable devices will clarify the association between indoor temperature and blood pressure
    Nakagami, Hironori
    Otsuka, Hiroki
    Akiyama, Hitoshi
    HYPERTENSION RESEARCH, 2023, 46 (06) : 1450 - 1455
  • [44] Monitoring with wearable devices will clarify the association between indoor temperature and blood pressure
    Hironori Nakagami
    Hiroki Otsuka
    Hitoshi Akiyama
    Hypertension Research, 2023, 46 : 1450 - 1455
  • [45] Blood pressure differences between northern and southern Chinese: Role of dietary factors the international study on macronutrients and blood pressure
    Zhao, LC
    Stamler, J
    Yan, LJL
    Zhou, BF
    Wu, YF
    Liu, K
    Daviglus, ML
    Dennis, BH
    Elliott, P
    Ueshima, H
    Yang, J
    Zhu, LG
    Guo, DS
    HYPERTENSION, 2004, 43 (06) : 1332 - 1337
  • [46] Comparison of two blood pressure oscillometric devices: Datascope Accutorr Plus and Mobil-O-Graph PWA and conversion of blood pressure values from one device to the other
    Sarganas, Giselle
    Kuhnert, Ronny
    Gohlisch, Christopher
    van der Giet, Markus
    Neuhauser, Hannelore
    BLOOD PRESSURE MONITORING, 2020, 25 (01) : 42 - 49
  • [47] Comparison of blood pressure measurements between an automated oscillometric device and a Hawksley random-zero sphygmomanometer in the northern Sweden MONICA study
    Eriksson, Marie
    Carlberg, Bo
    Jansson, Jan-Hakan
    BLOOD PRESSURE MONITORING, 2012, 17 (04) : 164 - 170
  • [48] Differences in Manual and Automatic Blood Pressures in Telemetry Patients With Atrial Fibrillation
    Collins, Melissa
    Cummings, Adam
    Skaggs, Casie
    Weller, Robyn
    Cronin, Sherill Nones
    DIMENSIONS OF CRITICAL CARE NURSING, 2013, 32 (05) : 262 - 265
  • [49] Wearable Wireless-Enabled Oscillometric Sphygmomanometer: A Flexible Ambulatory Tool for Blood Pressure Estimation
    Juteau, Nicolas
    Gosselin, Benoit
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2020, 14 (06) : 1287 - 1298
  • [50] Agreement of Doppler Ultrasound and Visual Sphygmomanometer Needle Oscillation with Invasive Blood Pressure in Anaesthetised Dogs
    Armour, Marc
    Michou, Joanne
    Schofield, Imogen
    Borland, Karla
    ANIMALS, 2024, 14 (19):