Collection efficiency of a monitor parallel plate ionization chamber for pencil beam scanning proton therapy

被引:0
作者
Rong-Cheng Han
Yong-Jiang Li
Yue-Hu Pu
机构
[1] Chinese Academy of Sciences,Shanghai Institute of Applied Physics
[2] University of Chinese Academy of Sciences,Shanghai Advanced Research Institute
[3] Shanghai APACTRON Particle Equipment Co. Ltd,undefined
[4] Chinese Academy of Sciences,undefined
来源
Nuclear Science and Techniques | 2020年 / 31卷
关键词
Collection efficiency; Monitor ionization chamber; Pencil beam scanning; Proton therapy;
D O I
暂无
中图分类号
学科分类号
摘要
The collection efficiency of monitor parallel plate ionization chambers is the main uncertainty in the beam control of pencil beam scanning systems. Existing calculation methods for collection efficiency in photon or passive scattering proton systems have not considered the characteristics of non-uniform charge density in pencil beam scanning systems. In this study, Boag’s theory was applied to a proton pencil beam scanning system. The transverse distribution of charge density in the ionization chamber was considered to be a Gaussian function and an analytical solution was derived to calculate collection efficiency in the beam spot area. This calculation method is called the integral method and it was used to investigate the effects of beam parameters on collection efficiency. It was determined that collection efficiency is positively correlated with applied voltage, beam size, and beam energy, but negatively correlated with beam current intensity. Additionally, it was confirmed that collection efficiency is improved when the air filling the monitor parallel plate ionization chamber is replaced with nitrogen.
引用
收藏
相关论文
共 81 条
  • [1] Wilson RR(1946)Radiological use of fast protons Radiology 47 487-491
  • [2] Lomax AJ(1999)Intensity modulation methods for proton radiotherapy Phys. Med. Biol. 44 185-205
  • [3] Zhao BQ(2017)The front-end electronics design of dose monitors for beam delivery system of Shanghai advanced proton therapy facility Nucl. Sci. Tech. 28 83-234
  • [4] Zhao MH(2017)Precise magnetic field control of the scanning magnets for the APTRON beam delivery system Nucl. Sci. Tech. 28 172-e1072
  • [5] Liu M(2016)Simulation of spot scanning in proton therapy Nucl. Tech. 39 090202-6880
  • [6] Miao C(2010)Dosimetry for ion beam radiotherapy Phys. Med. Biol. 55 193-102
  • [7] Liu M(2018)Dose detectors, sensors, and their applications Med. Phys. 45 e1051-1870
  • [8] Yin C(2015)Dosimetric accuracy of a treatment planning system for actively scanned proton beams and small target volumes: Monte Carlo and experimental validation Phys. Med. Biol. 60 6865-344
  • [9] Jia YJ(2016)EPR/alanine dosimetry for two therapeutic proton beams Nucl. Instrum. Meth. B 368 96-126
  • [10] Li YJ(1999)AAPM’s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Med. Phys. 26 1847-3558