Biomechanical analysis of a radial functionally graded dental implant–bone system under multi-directional dynamic loads

被引:0
作者
A. Jarrahi
H. Asgharzadeh Shirazi
A. Asnafi
M. R. Ayatollahi
机构
[1] Iran University of Science and Technology,Tissue Engineering and Biological Systems Research Laboratory, School of Mechanical Engineering
[2] Iran University of Science and Technology,Fatigue and Fracture Research Laboratory, Center of Excellence in Experimental Solid Mechanics and Dynamics, School of Mechanical Engineering
[3] Shiraz University,Hydro
来源
Journal of the Brazilian Society of Mechanical Sciences and Engineering | 2018年 / 40卷
关键词
Dental implant system; Functionally graded biomaterials; Dynamic behavior; Finite element analysis;
D O I
暂无
中图分类号
学科分类号
摘要
Development of new biomaterials such as FGBMs for dental clinical implant applications have been recently attracted a lot of attention due to their advantages and abilities in satisfying both mechanical and biocompatibility properties simultaneously. Drawing inspiration from the geometry of a real dental implant system, and using the properties of radial functionally graded biomaterials, a model was investigated to fabricate the artificial FGBM dental implant under periodic dynamic loads. The investigation is done using a finite element analyzer, and the problem is solved under the actual bone geometry, constraints, conditions and combined dynamic loads. The effects of radial FGBM dental implants on the stress and deformation values near the dental implant–bone interface under periodic dynamic loads are simulated. Response characteristics of FGBM implant models with different heterogeneous biomaterial parameters are determined and compared to the behavior of conventional titanium and titanium coated by a film of hydroxyapatite implants. The results show that using radial FGBM dental implants can improve the mechanical behaviors of dental implant systems regarding stress and deformation. It was also found that FGBM parameters can play an essential role in achieving an efficient biomechanical function in those systems.
引用
收藏
相关论文
共 140 条
  • [21] Kjellin P(2012)Modification of electrophoretically deposited nano-hydroxyapatite coatings by wire brushing on Ti–6Al–4V substrates Ceram Int 38 4885-4893
  • [22] Wennerberg A(2005)Hydroxyapatite-coated metals: interfacial reactions during sintering J Mater Sci 16 101-106
  • [23] Elias CN(2011)Induction plasma sprayed nano hydroxyapatite coatings on titanium for orthopaedic and dental implants Surf Coat Technol 205 2785-2792
  • [24] Rocha FA(2005)Design of functionally graded dental implant in the presence of cancellous bone J Biomed Mater Res Part B 75 74-80
  • [25] Nascimento AL(2007)Thermal-mechanical study of functionally graded dental implants with the finite element method J Biomed Mater Res Part A 80 146-158
  • [26] Coelho PG(2007)A three-dimensional finite element study on the biomechanical behavior of an FGBM dental implant in surrounding bone J Biomech 40 2377-2385
  • [27] Demenko V(2006)Static, dynamic and fatigue behaviors of dental implant using finite element method Adv Eng Softw 37 649-658
  • [28] Linetskiy I(2010)Analysis of the effect of load direction on the stress distribution in the dental implant Mater Des 31 2097-2101
  • [29] Nesvit K(2007)Restoration of non-carious cervical lesions: part II. Restorative material selection to minimise fracture Dent Mater 23 1562-1569
  • [30] Hubalkova H(2007)Functional significance of strain distribution in the human mandible under masticatory load: numerical predictions Arch Oral Biol 52 465-473