Fracture Outer Metallic Tracheostomy Tube as an Airway Foreign Body

被引:0
作者
Patnarin Mahattanasakul
Tonphong Kaewkongka
Thitiwat Sriprasart
Virachai Kerekhanjanarong
机构
[1] King Chulalongkorn Memorial Hospital,Department of Otolaryngology, Head and Neck Surgery
[2] Thai Red Cross Society,Department of Otolaryngology, Head and Neck Surgery, Faculty of Medicine
[3] Chulalongkorn University,Department of Physics, Faculty of Sciences
[4] Chulalongkorn University,Division of Pulmonary and Critical Care Medicine, Department of Medicine, Faculty of Medicine
[5] Chulalongkorn University,Division of Pulmonary and Critical Care Medicine, Department of Medicine
[6] King Chulalongkorn Memorial Hospital,undefined
[7] Thai Red Cross Society,undefined
来源
Indian Journal of Otolaryngology and Head & Neck Surgery | 2022年 / 74卷
关键词
Foreign bodies; Fracture; Respiratory aspiration; Tracheostomy;
D O I
暂无
中图分类号
学科分类号
摘要
To study the clinical presentation, management and mechanism of fractured outer metallic tracheostomy tube presenting as tracheobronchial foreign body. A retrospective chart review patients with fracture outer metallic tracheostomy tube. Data regarding the patients’ demographic data, diagnosis, clinical presentation, type of tracheostomy tube and site of fracture were analyzed. Total 4 cases of fracture outer metallic tracheostomy tube were found. There were 3 males and 1 female, average age 52.75 years, range 31–76 years. The common presentation were dyspnea, intolerable cough and decreased breath sound in 4(100%), 2(50%) and 2(50%) cases. The most serious presentation was cardiac arrested 1 case. The dislodged tube were retrieved by flexible and rigid bronchoscopy. The most common site of fracture were outer tube at mid shaft 3 cases (75%). All of this site had corrosion. Only 1 case (25%) was fracture at junction between neck plate and tube without corrosion. The average time of usage metallic tracheostomy tube was 24 days, range 3–34 days. Fracture tracheostomy tube is rare and serious medical emergency. The patients, caregivers and physicians should recognition and prompt action. Flexible or rigid bronchoscopy via tracheostoma can successfully removal the dislodge part. The mechanism of fracture may come from several factors. The defective manufacturer, stagnation of alkaline bronchial secretion, recurrence process of removal, cleaning and boiling of the tube can cause mechanical stress and degradation of passive film of the tubes. The patient education regarding the maintenance and regular checked up can possibly extinguish this complication.
引用
收藏
页码:1752 / 1756
页数:4
相关论文
共 59 条
  • [1] Stauffer JL(1981)Complication and consequences of endotracheal tube intubation and tracheotomy. A prospective study of 150 critically ill adult patients Am J Med 70 65-76
  • [2] Olson DE(2010)Fracture metallic tracheosotmy tube in a child: a case report and review of the literature J Med Case Rep 4 234-237
  • [3] Petty TL(2009)Fracture and aspiration of tracheostomy tube Tanaffos 8 75-78
  • [4] Piromchai P(1960)Broken tracheotomy tube as a foreign body Lancet 1 1006-1007
  • [5] Lertchanaruengrit P(1972)Inner migration of the inner cannula: an unusual foreign body Eye Ear Nose Throat Mon 51 257-258
  • [6] Vatansapt P(1994)Fracture of a synthetic fenestrated tracheotomy tube: case report and review of the literature Am J Otolaryngol 15 63-67
  • [7] Ratanaanekchai T(1990)Corrosion and fracture of a silver tracheotomy tube J Laryngol Otol 104 48-49
  • [8] Thanaviratananich S(1989)Fractured tracheotomy tube in the tracheobronchial tree J Laryngol Otol 103 335-336
  • [9] Radpay B(2014)Fractured and migrated tracheotomy tube in the tracheobronchial tree Int J Pediatr Otorhinolaryngol 78 1472-1475
  • [10] Pejhan S(1993)Broken tracheostomy introducer—an unusual tracheobronchial foreign body J Laryngol Otol 107 463-464