Improved sample preparation procedure for gross alpha counting on gas-flow proportional counter

被引:0
|
作者
Adam G. Burn
Xin Li
Douglas K. Haines
Mirza M. Hussain
Karl X. Yang
Joseph C. Marrantino
Umme-Farzana Syed
Abdul Bari
Thomas M. Semkow
机构
[1] New York State Department of Health,Wadsworth Center
[2] University of Albany,Department of Environmental Health Sciences, School of Public Health
[3] State University of New York,undefined
关键词
Gross alpha activity; Drinking water; Self-absorption; Evaporation; Coprecipitation; Power law;
D O I
暂无
中图分类号
学科分类号
摘要
Sample residue uniformity is crucial for accurate gross alpha counting using a gas-flow proportional counter, however this can be difficult to achieve using conventional procedures. An evaporation method using a 1:1 mixture of deionized water and 2-propanol yielded uniform 230Th and 241Am nitrate salt residues on metal planchets for gross alpha counting. The gross alpha efficiency and alpha:beta crosstalk behavior were studied using the evaporation procedure and a separate barium sulfate/ferric hydroxide coprecipitation method. The gross alpha efficiency data were fitted with empirical power law functions. The alpha:beta crosstalk behavior was different for the two radionuclides studied.
引用
收藏
页码:417 / 423
页数:6
相关论文
共 50 条
  • [41] Improving accuracy and precision in gross alpha and beta counting with proportional detectors
    Abrantes, J.
    Pinhao, N. R.
    Melo, J.
    Madruga, M. J.
    NUCLEAR PROFICIENCY TESTING, 2008, 1036 : 186 - 200
  • [42] ABSOLUTE BETA-COUNTING IN A PROPORTIONAL FLOW-COUNTER
    ROALSVIG, JP
    HASLAM, RNH
    CANADIAN JOURNAL OF PHYSICS, 1959, 37 (04) : 499 - 514
  • [43] OPTIMIZATION OF A CONVERSION ELECTRON MOSSBAUER-SPECTROSCOPY GAS-FLOW HE/CH4 PROPORTIONAL COUNTER
    HANZEL, D
    GRIESBACH, P
    MEISEL, W
    GUTLICH, P
    HYPERFINE INTERACTIONS, 1992, 71 (1-4): : 1441 - 1444
  • [44] Calibration of a low background gas-flow proportional counter to estimate 234Th activity in coastal waters
    Cuesta, E.
    Lozano, R. L.
    San Miguel, E. G.
    Casas-Ruiz, M.
    Bolivar, J. P.
    APPLIED RADIATION AND ISOTOPES, 2016, 118 : 201 - 210
  • [45] ESTIMATION OF THE ENERGY-DISTRIBUTION OF MOSSBAUER SCATTERED ELECTRONS WITH 2-PI GAS-FLOW PROPORTIONAL COUNTER
    MATSUO, M
    SATO, H
    TOMINAGA, T
    RADIOCHEMICAL AND RADIOANALYTICAL LETTERS, 1981, 48 (05): : 253 - 262
  • [46] THE USE OF A GAS-FLOW PROPORTIONAL COUNTER FOR THE ENERGETICAL ANALYSIS OF PHOTOELECTRONS UNDER X-RAY-DIFFRACTION CONDITIONS
    KOVALCHUK, MV
    MUKHAMEDZHANOV, EK
    PHYSICA STATUS SOLIDI A-APPLIED RESEARCH, 1984, 81 (02): : 427 - 432
  • [47] A GAS-FLOW X-RAY DIFFRACTION COUNTER
    ARNDT, UW
    COATES, WA
    CRATHORN, AR
    PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON SECTION B, 1954, 67 (412): : 357 - 359
  • [48] A GAS-FLOW PROPORTIONAL COUNTER FOR THE MOSSBAUER STUDY OF SURFACES BETWEEN 100-K AND 400-K
    COOK, DC
    HYPERFINE INTERACTIONS, 1986, 29 (1-4): : 1463 - 1466
  • [49] Determination of radon in water by argon purging and alpha counting with a proportional counter
    Zikovsky, L.
    Roireau, N.
    Applied Radiation and Isotopes, 1990, 41 (07): : 679 - 681
  • [50] A NEW WINDOW FLOW PROPORTIONAL COUNTER FOR COUNTING RADIOACTIVE GAS-LIQUID PARTITION CHROMATOGRAPHY EFFLUENTS
    WELCH, M
    WITHNELL, R
    WOLF, AP
    ANALYTICAL CHEMISTRY, 1967, 39 (02) : 275 - &