Sensitive and selective determination of mercury by differential pulse stripping voltammetry after accumulation of mercury vapour on a gold plated graphite electrode

被引:0
|
作者
Mieczysław Korolczuk
机构
[1] Maria Curie Skłodowska University,Faculty of Chemistry
来源
Fresenius' Journal of Analytical Chemistry | 1997年 / 357卷
关键词
Mercury; Differential Pulse; Mercury Vapour; Anal Chim; Anodic Strip Voltammetry;
D O I
暂无
中图分类号
学科分类号
摘要
A selective and sensitive method is proposed for the determination of mercury by anodic stripping voltammetry after its preconcentration from the gas phase. Mercury from the sample solution is reduced to elemental Hg by SnCl2 and volatilized by the bubbles of a carrier gas. The gas containing mercury vapour is dried and passed through a capillary onto a gold coated graphite electrode. An anodic stripping voltam-mogram is recorded from 0.1 mol/1 HClO4 + 3 × 10−3 mol/1 HCl solution. The calibration curve is linear from 1 × 10−9 to 4 × 10−8 mol/1 Hg(NO3)2. The absolute detection limit is 0.46 ng Hg. The relative standard deviations for 4 × 10−9 mol/1 and 2 × 10−8 mol/1 Hg(NO3)2 are 9.8% and 6.1%, respectively n = 5).
引用
收藏
页码:389 / 391
页数:2
相关论文
共 50 条
  • [31] Imprinted polymer-modified hanging mercury drop electrode for differential pulse cathodic stripping voltammetric analysis of creatine
    Lakshmi, Dhana
    Sharma, Piyush S.
    Prasad, Bhim B.
    BIOSENSORS & BIOELECTRONICS, 2007, 22 (12) : 3302 - 3308
  • [32] The Graphene/L-Cysteine/Gold-Modified Electrode for the Differential Pulse Stripping Voltammetry Detection of Trace Levels of Cadmium
    Song, Yu
    Bian, Chao
    Tong, Jianhua
    Li, Yang
    Xia, Shanghong
    MICROMACHINES, 2016, 7 (06):
  • [33] Determination of Nanomolar Mercury(II) Concentration by Anodic-Stripping Voltammetry at a Carbon Paste Electrode Modified with Functionalized Nanoporous Silica Gel
    Javanbakht, Mehran
    Khoshsafar, Hamid
    Ganjali, Mohammad R.
    Badiei, Alireza
    Norouzi, Parviz
    Hasheminasab, Abed
    CURRENT ANALYTICAL CHEMISTRY, 2009, 5 (01) : 35 - 41
  • [34] POTENTIOMETRIC STRIPPING DETERMINATION OF MERCURY(II), SELENIUM(IV), COPPER(II) AND LEAD(II) AT A GOLD FILM ELECTRODE IN WATER SAMPLES
    GIL, EP
    OSTAPCZUK, P
    ANALYTICA CHIMICA ACTA, 1994, 293 (1-2) : 55 - 65
  • [35] Poly(3,4-ethylenedioxythiophene)/Electrochemically Reduced Graphene Oxide Film Modified Electrode for Mercury(II) Determination Using Anodic Stripping Voltammetry
    Kalaivani, A.
    Saranya, N.
    Divya, R.
    Narayanan, S. Sriman
    GRAPHENE, 2013, 1 (02) : 131 - 135
  • [36] Determination of Ultra-trace Mercury(II) by Flow-injection/Anodic Stripping Voltammetry Using a Track-etched Microporous Membrane Electrode
    Mizuguchi, Hitoshi
    Numata, Kentaro
    Monma, Chiaki
    Iiyama, Masamitsu
    Tachibana, Kazuhiro
    Nishina, Tatsuo
    Shida, Junichi
    ANALYTICAL SCIENCES, 2013, 29 (10) : 949 - 954
  • [37] Determination of Ultra-trace Mercury(II) by Flow-injection/ Anodic Stripping Voltammetry Using a Track-etched Microporous Membrane Electrode
    Hitoshi Mizuguchi
    Kentaro Numata
    Chiaki Monma
    Masamitsu Iiyama
    Kazuhiro Tachibana
    Tatsuo Nishina
    Junichi Shida
    Analytical Sciences, 2013, 29 : 949 - 954
  • [38] Continuous flow and flow injection stripping voltammetric determination of silver(I), mercury(II), and bismuth(III) at a bulk modified graphite tube electrode
    Ye, RD
    Khoo, SB
    ELECTROANALYSIS, 1997, 9 (06) : 481 - 489
  • [39] Analytical procedure for total mercury determination in fishes and shrimps by chronopotentiometric stripping analysis at gold film electrodes after microwave digestion
    Augelli, M. A.
    Munoz, R. A. A.
    Richter, E. M.
    Cantagallo, M. I.
    Angnes, L.
    FOOD CHEMISTRY, 2007, 101 (02) : 579 - 584
  • [40] ELECTROCHEMICAL DETERMINATION OF ULTRALOW LEVELS (LESS-THAN-10(-12)-M) OF MERCURY BY ANODIC-STRIPPING VOLTAMMETRY USING CHEMICALLY-MODIFIED ELECTRODE
    TURYAN, I