Flexible impeller blades in an axial flow pump for intravascular cavopulmonary assistance of the fontan physiology

被引:0
|
作者
Throckmorton A.L. [1 ]
Kapadia J.Y. [1 ]
Carr J.P. [1 ]
Powell C.M. [1 ]
Tate R.D. [1 ]
Traynham D.V. [1 ]
机构
[1] Department of Mechanical Engineering, BioCirc Research Laboratory, Virginia Commonwealth University, Richmond, VA 23284, 401 West Main Street, E3221
基金
美国国家科学基金会;
关键词
Artificial right ventricle; Blood pump; Cavopulmonary assist device; Heart pump; Intravascular blood pump; Mechanical cavopulmonary assist; Pediatric circulatory support; Single ventricle physiology;
D O I
10.1007/s13239-010-0026-2
中图分类号
学科分类号
摘要
An intravascular axial flow blood pump has suitable hydraulic characteristics for use to augment pressure in the cavopulmonary circulation in Fontan patients. This study presents the experimental hydraulic performance testing of six flexible-bladed and one rigid-bladed pump prototypes for mechanical cavopulmonary assist. Unique in design, six of these prototypes were manufactured using varying grades of polyurethane material to reduce blade hardness. The hydraulic performance of these prototypes were measured and quantitatively compared to a rigid-bladed prototype through a regression analysis. To shield the vessel wall from the rotating impeller blades, two novel protective cage designs were also evaluated for additional energy augmentation. The pumps with no protective cage produced pressure rises of 1-12 mmHg for flow rates of 0.5-4 L/min at rotational speeds of 4000-7000 rpm. The flexible-bladed prototypes generated pressure rises within 12% of the rigid-bladed pump design. The most flexible prototype (F-15) was able to generate pressure rises within 4.1% of the rigid-bladed impeller. The partially twisted and fully twisted protective cage designs also functioned as desired by augmenting energy transfer in range of 5-15 mmHg, respectively. This study dispels the assumption that rigid-impeller blades are absolutely necessary to maintain the pressure generation of rotary blood pumps and demonstrates the potential of uniquely shaped cage filaments to augment the energy transfer of an axial flow blood pump for mechanical cavopulmonary assistance. The successful development of this blood pump will provide clinicians with a therapeutic option for mechanically supporting the failing Fontan physiology in adolescent and adult patients. © 2010 Biomedical Engineering Society.
引用
收藏
页码:244 / 255
页数:11
相关论文
共 50 条
  • [1] Intravascular Mechanical Cavopulmonary Assistance for Patients With Failing Fontan Physiology
    Bhavsar, Sonya S.
    Kapadia, Jugal Y.
    Chopski, Steven G.
    Throckmorton, Amy L.
    ARTIFICIAL ORGANS, 2009, 33 (11) : 977 - 987
  • [2] MECHANICAL ASSISTANCE FOR THE FONTAN CIRCULATION USING AN INTRAVASCULAR AXIAL FLOW BLOOD PUMP
    Bhavsar, Sonya S.
    Moskowitz, William B.
    Throckmorton, Amy L.
    PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE, 2010, 2010, : 615 - 616
  • [3] Shape Optimization of Impeller Blades for Bidirectional Axial Flow Pump
    Baek, Seok Heum
    Jung, Won Hyuk
    Kang, Sangmo
    TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS B, 2012, 36 (12) : 1141 - 1150
  • [4] Cavopulmonary Support with a Microaxial Pump for the Failing Fontan Physiology
    Zhu, Jiaquan
    Kato, Hideyuki
    Fu, Yaqin Y.
    Zhao, Lisa
    Foreman, Celeste
    Davey, Lisa
    Weisel, Richard D.
    Van Arsdell, Glen S.
    Honjo, Osami
    ASAIO JOURNAL, 2015, 61 (01) : 49 - 54
  • [5] Performance evaluation of a pediatric viscous impeller pump for Fontan cavopulmonary assist
    Giridharan, Guruprasad A.
    Koenig, Steven C.
    Kennington, Jeffrey
    Sobieski, Michael A.
    Chen, Jun
    Frankel, Steven H.
    Rodefeld, Mark D.
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 145 (01): : 249 - 257
  • [6] Cavopulmonary assist for the univentricular Fontan circulation: von Karman viscous impeller pump
    Rodefeld, Mark D.
    Coats, Brandon
    Fisher, Travis
    Giridharan, Guruprasad A.
    Chen, Jun
    Brown, John W.
    Frankel, Steven H.
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 140 (03): : 529 - 536
  • [7] Twisted cardiovascular cages for intravascular axial flow blood pumps to support the Fontan physiology
    Throckmorton, Amy L.
    Downs, Emily A.
    Hazelwood, John A.
    Monroe, Jonathan O.
    Chopski, Steven G.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2012, 35 (05): : 369 - 375
  • [8] Viscous Impeller Pump (VIP™) for Fontan Cavopulmonary Assist: Performance at Neonate and Infant scale
    Rodefeld, Mark
    Kennington, Jeff
    Giridharan, Guruprasad
    Koenig, Steven
    Jun, Chen
    Frankel, Steven
    CIRCULATION, 2010, 122 (21)
  • [9] Controlled Pitch-Adjustment of Impeller Blades for an Intravascular Blood Pump
    Throckmorton, Amy L.
    Sciolino, Michael G.
    Downs, Emily A.
    Saxman, Robert S.
    Lopez-Isaza, Sergio
    Moskowitz, William B.
    ASAIO JOURNAL, 2012, 58 (04) : 382 - 389
  • [10] Design of a Protective Cage for an Intravascular Axial Flow Blood Pump to Mechanically Assist the Failing Fontan
    Throckmorton, Amy L.
    Kishore, Ravi A.
    ARTIFICIAL ORGANS, 2009, 33 (08) : 611 - 621