The evaluation and calibration of fan-beam collimators

被引:0
|
作者
Jennifer L. Mahowald
Peter D. Robins
Michael K. O’Connor
机构
[1] Department of Radiology,
[2] Mayo Clinic,undefined
[3] Rochester,undefined
[4] MN 55905,undefined
[5] USA,undefined
来源
European Journal of Nuclear Medicine | 1999年 / 26卷
关键词
Key words: Fan-beam collimator; Single-photon emission tomography; Quality control;
D O I
暂无
中图分类号
学科分类号
摘要
The aims of this study were (a) to determine the true focal length of a fan-beam collimator and (b) to calibrate image size (mm/pixel) for each collimator to permit inter-comparison of image data acquired on different gamma camera systems. A total of six fan-beam collimators on three dual-head gamma camera systems were evaluated using a set of four cobalt-57 point source markers. The markers were arranged in a line in the transverse plane with a known separation between them. Tomographic images were obtained at three radii of rotation. From reconstructed transaxial images the distance between markers was measured in pixels and used to determine pixel size in mm/pixel. The system value for the focal length of the collimator was modified by up to ±100 mm and transaxial images were again reconstructed. To standardize pixel size between systems, the apparent radius of rotation during a single-photon emission tomography (SPET) acquisition was modified by changes to the effective collimator thickness. SPET images of a 3D brain phantom were acquired on each system and reconstructed using both the original and the modified values of collimator focal length and thickness. Co-registration and subtraction of the reconstructed transaxial images was used to evaluate the effects of changes in collimator parameters. Pixel size in the reconstructed image was found to be a function of both the radius of rotation and the focal length. At the correct focal length, pixel size was essentially independent of the radius of rotation. For all six collimators, true focal length differed from the original focal length by up to 26 mm. These differences in focal length resulted in up to 6% variation in pixel size between systems. Pixel size between the three systems was standardized by altering the value for collimator thickness. Subtraction of the co-registered SPET images of the 3D brain phantom was significantly improved after optimization of collimator parameters, with a 35%–50% reduction in the standard deviation of residual counts in the subtraction images. In conclusion, we have described a simple method for measurement of the focal length of a fan-beam collimator. This is an important parameter on multidetector systems for optimum image quality and where accurate co-registration of SPET to SPET and SPET to MRI studies is required.
引用
收藏
页码:314 / 319
页数:5
相关论文
共 18 条
  • [1] The evaluation and calibration of fan-beam collimators
    Mahowald, JL
    Robins, PD
    O'Connor, MK
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (04) : 314 - 319
  • [2] Characterisation of fan-beam collimators
    Pareto, D
    Pavía, J
    Falcón, C
    Juvells, I
    Cot, A
    Ros, D
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2001, 28 (02) : 144 - 149
  • [3] The role of AGFA high-energy CR in the calibration and quality control of multileaf collimators (MLC)
    Li Zhaobin
    Xiong Fei
    Huang Guofeng
    Cao Zheng
    Jiang Ruiyao
    Fu Shen
    NUCLEAR SCIENCE AND TECHNIQUES, 2007, 18 (03) : 145 - 149
  • [4] The role of AGFA high-energy CR in the calibration and quality con-trol of multileaf collimators(MLC)
    LI Zhaobin XIONG Fei HUANG Guofeng CAO Zheng JIANG Ruiyao FU Shen* (Department of Radiation Oncology
    Nuclear Science and Techniques, 2007, (03) : 145 - 149
  • [5] Evaluation of optimally designed planar-concave collimators in single-photon emission tomography
    Kimiaei, S
    Ljungberg, M
    Larsson, SA
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1997, 24 (11) : 1398 - 1404
  • [6] Evaluation of optimally designed planar-concave collimators in single-photon emission tomography
    Shahrok Kimiaei
    Michael Ljungberg
    Stig A. Larsson
    European Journal of Nuclear Medicine, 1997, 24 : 1398 - 1404
  • [7] Evaluation of the planar sources surface homogeneity used to instruments calibration
    Vivolo, Vitor
    Albuquerque Potiens, Maria da Penha
    APPLIED RADIATION AND ISOTOPES, 2010, 68 (4-5) : 605 - 606
  • [8] Two quality control procedures on radiotherapy beam calibration and treatment planning system implementation
    Piermattei, A
    Azario, L
    delle Canne, S
    Fidanzio, A
    Palazzoni, G
    TUMORI, 1998, 84 (02) : 150 - 154
  • [9] Evaluation of in situ sea surface temperatures for use in the calibration and validation of satellite retrievals
    Xu, Feng
    Ignatov, Alexander
    JOURNAL OF GEOPHYSICAL RESEARCH-OCEANS, 2010, 115
  • [10] Evaluation of on-site calibration procedures for SKYNET Prede POM sun-sky photometers
    Campanelli, Monica
    Estelles, Victor
    Kumar, Gaurav
    Nakajima, Teruyuki
    Momoi, Masahiro
    Grobner, Julian
    Kazadzis, Stelios
    Kouremeti, Natalia
    Karanikolas, Angelos
    Barreto, Africa
    Nevas, Saulius
    Schwind, Kerstin
    Schneider, Philipp
    Harju, Iiro
    Karha, Petri
    Diemoz, Henri
    Kudo, Rei
    Uchiyama, Akihiro
    Yamazaki, Akihiro
    Iannarelli, Anna Maria
    Mevi, Gabriele
    Di Bernardino, Annalisa
    Casadio, Stefano
    ATMOSPHERIC MEASUREMENT TECHNIQUES, 2024, 17 (17) : 5029 - 5050