Consequences of high-sensitivity troponin T testing applied in a primary care population with chest pain compared with a commercially available point-of-care troponin T analysis: An observational prospective study Cardiovascular Disorders

被引:9
作者
Andersson P.O. [1 ]
Karlsson J.-E. [2 ,3 ]
Landberg E. [4 ]
Festin K. [3 ]
Nilsson S. [5 ]
机构
[1] Primary Health Care and Department of Medical and Health Sciences, Linköping University, Linköping
[2] Department of Internal Medicine, County Council of Jönköping, Jönköping
[3] Department of Medical and Health Sciences, Linköping University, Linköping
[4] Department of Clinical Chemistry, Linköping University, Linköping
[5] Primary Health Care, Department of Medical and Health Sciences, Linköping University, Norrköping
关键词
Acute coronary syndrome; Chest pain; Point-of-care testing; Primary health care; Troponin T;
D O I
10.1186/s13104-015-1174-0
中图分类号
学科分类号
摘要
Background: There is a demand for a highly sensitive and specific point-of care test to detect acute myocardial infarction (AMI). It is unclear if a high-sensitivity troponin assay will have enough discriminative power to become a decision support in primary care. The aim of this study was to evaluate a high-sensitivity troponin T assay performed in three primary health care centres in southeast Sweden and to compare the outcome with a point-of-care troponin T test. Methods: This study included 115 patients who consulted their general practitioner for chest pain, dyspnoea on exertion, unexplained weakness and/or fatigue in the last 7 days. Troponin T was analysed by a point-of-care test and a high-sensitivity method together with N-terminal pro-B-type natriuretic peptide (NT-proBNP) and creatinine. All patients were checked for AMI or unstable angina (UA) within 30 days of study enrolment. Univariate and multivariate logistic regression was carried out to examine possible connections between troponin T ≥ 15 ng/L, clinical variables and laboratory findings at baseline. In addition, 21 patients with troponin T ≥ 15 ng/L and no signs of AMI or UA were followed up for 2-3 years. Results: Three patients were diagnosed with AMI and three with UA. At the ≥ 15 ng/L cut-off, the troponin T method had 100% sensitivity, 75% specificity for AMI and a positive predictive value of 10%. The troponin T point-of-care test missed one case of AMI and the detection limit was 50 ng/L. Troponin T ≥ 15 ng/L was correlated to age ≥65 years (odds ratio (OR), 10.9 95% CI 2.28-51.8) and NT-proBNP in accordance with heart failure (OR 8.62 95% CI 1.61-46.1). Fourteen of the 21 patients, without signs of AMI or UA at baseline, still had increased troponin T at follow-up after 2-3 years. Conclusions: A high-sensitivity troponin T assay could become useful in primary care as a point-of-care test for patients <65 years. For patients older than 65-70 years, a higher decision limit than ≥15 ng/L should be considered and used in conjunction with clinical parameters and possibly with NT-proBNP. © 2015 Andersson et al.
引用
收藏
相关论文
共 19 条
  • [1] Cost consequences of point-of-care troponin T testing in a Swedish primary health care setting
    Nilsson, Staffan
    Andersson, Agneta
    Janzon, Magnus
    Karlsson, Jan-Erik
    Levin, Lars-Ake
    SCANDINAVIAN JOURNAL OF PRIMARY HEALTH CARE, 2014, 32 (04) : 241 - 247
  • [2] Survival and resource utilization in patients with chest pain evaluated with cardiac troponin T compared with high-sensitivity cardiac troponin T
    Bandstein, Nadia
    Wikman, Anna
    Ljung, Rickard
    Holzmann, Martin J.
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2017, 245 : 43 - 48
  • [3] Point-of-Care High-Sensitivity Troponin Testing in the Emergency Department The Way of the Future?
    Clerico, Aldo
    Aimo, Alberto
    Passino, Claudio
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 84 (08) : 741 - 743
  • [4] Diagnostic performance of a point-of-care high-sensitivity cardiac troponin I assay among Chinese patients with chest pain
    Ho, Jonathan Ka Ming
    Wong, Janet Yuen Ha
    Tse, Gary
    Chong, Andy Chun Yin
    Chau, Calvin Chi Wai
    Wong, Chi Yip
    Tse, Johnson Wai Keung
    Tam, Jeremy Yan Hon
    Lam, Simon Ching
    OPEN HEART, 2024, 11 (02):
  • [5] Diagnostic Performance of Point-of-Care Troponin I and Laboratory Troponin T in Patients Presenting to the ED with Chest Pain: A Comparative Study
    Mohammad, Osama H.
    Naushad, Vamanjore A.
    Purayil, Nishan K.
    Sinan, Laith
    Ambra, Naseem
    Chandra, Prem
    Paramba, Firjeeth C.
    Mohammad, Jassim
    Chalihadan, Sajid
    Varikkodan, Irfan
    Palol, Azeez
    OPEN ACCESS EMERGENCY MEDICINE, 2020, 12 : 247 - 254
  • [6] Risk stratification of chest pain patients by point-of-care cardiac troponin T and myoglobin measured in the emergency department
    Ordónez-Llanos, J
    Santalò-Bel, M
    Mercé-Muntañola, J
    Collinson, PO
    Gaze, D
    Haass, M
    Katus, HA
    Chwallek, F
    Hirschl, MM
    Derhaschnig, U
    Mueller-Bardorff, M
    Kellett, J
    Sylvén, C
    Schulz, I
    Zerback, R
    CLINICA CHIMICA ACTA, 2006, 365 (1-2) : 93 - 97
  • [7] Study protocol for an observational study to evaluate an accelerated chest pain pathway using point-of-care troponin in New Zealand rural and primary care populations
    Miller, Rory
    Young, Joanna
    Nixon, Garry
    Pickering, John W.
    Stokes, Tim
    Turner, Robin
    Devlin, Gerard
    Watson, Antony
    Gutenstein, Marc
    Norman, Tim
    George, Peter Myles
    Du Toit, Stephen
    Than, Martin
    JOURNAL OF PRIMARY HEALTH CARE, 2020, 12 (02) : 129 - 138
  • [8] High-sensitivity troponin testing at the point of care for the diagnosis of myocardial infarction: a prospective emergency department clinical evaluation
    Curran, Jemima May
    Mergo, Aleksandra
    White, Sarah
    Croal, Bernard L.
    Cooper, Jamie G.
    EMERGENCY MEDICINE JOURNAL, 2024, : 320 - 321
  • [9] High-sensitivity troponin T testing and coronary computed tomography angiography for rapid diagnosis of chest pain in the emergency department
    Duran-Cambra, Albert
    Rossello, Xavier
    Sans-Rosello, Jordi
    Vila, Montserrat
    Hidalgo, Alberto
    Diaz-Rodriguez, Ivan
    Leta, Ruben
    Pons-Llado, Guillem
    Ordonez-Llanos, Jordi
    Sionis, Alessandro
    EMERGENCIAS, 2016, 28 (01): : 9 - 15
  • [10] Analytical comparison of the new point-of-care troponin T immunoassay on AQT90Flex® analyzer (Radiometer) and the high-sensitivity troponin T immunoassay on ModularE170® (Roche Diagnostics)
    Monneret, Denis
    Ben Abbes, Rachel Ahmed
    Omarjee, Riyadh
    Devilliers, Catherine
    Le Manach, Yannick
    Raux, Mathieu
    Coriat, Pierre
    Imbert-Bismut, Francoise
    Bonnefont-Rousselot, Dominique
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2014, 52 (12) : E279 - E282