A biomechanical study of plate versus intramedullary devices for midshaft clavicle fixation

被引:65
作者
Golish S.R. [1 ]
Oliviero J.A. [1 ]
Francke E.I. [1 ]
Miller M.D. [1 ]
机构
[1] Department of Orthopaedic Surgery, University of Virginia Health System, Charlottesville, VA
关键词
Intramedullary Fixation; Dynamic Compression Plate; Intramedullary Device; Plate Construct; Midshaft Clavicle Fracture;
D O I
10.1186/1749-799X-3-28
中图分类号
学科分类号
摘要
Background. Non-operative management of midshaft clavicle fractures is standard; however, surgical management is increasing. The purpose of this study is to compare the biomechanical performance of plate versus intramedullary fixation in cyclic bending for matched pairs of cadaveric clavicles. We hypothesized that the biomechanical properties are similar. Methods. Eight sets of matched clavicles with vertical, midshaft osteotomies were prepared from fresh, frozen cadavers. A 3.5 mm dynamic compression plate or a 3.8 or 4.5 mm intramedullary device were used for fixation. Clavicles were loaded in a four-point bend at 6 different loads for 3000 cycles at 1 Hz starting with 180 N and increasing by 180 N with sampling at 2 Hz. Failure was defined as 10 mm of displacement or catastrophic construct failure prior to 10 mm of displacement. Results. Between constructs, there was a significant difference with large effect size in displacement at fixed loads of 180 N (P = 0.001; Cohen's d = 1.85), 360 N (P = 0.033; Cohen's d = 1.39), 540 N (P = 0.003; Cohen's d = 0.73) and 720 N (P = 0.018; Cohen's d = 0.72). There was a significant difference with large effect size in load at fixed displacements of 5 mm (P = 0.001; Cohen's d = 1.49), 7.5 mm (P = 0.011; Cohen's d = 1.06), and 10 mm (P = 0.026; Cohen's d = 0.84). Conclusion. Plate constructs are superior in showing less displacement at fixed loads and greater loads at fixed displacements over a broad range of loads and displacements with cyclic four-point bending. The clinical relevance is that plate fixation may provide a stronger construct for early rehabilitation protocols that focus on repetitive movements in the early pre-operative period. © 2008 Golish et al; licensee BioMed Central Ltd.
引用
收藏
相关论文
共 12 条
[1]  
Robinson C.M., Fractures of the clavicle in the adult. Epidemiology and classification, J Bone Joint Surg Br, 80, 3, pp. 476-484, (1998)
[2]  
Stanley D., Trowbridge E.A., Norris S.H., The mechanism of clavicular fracture. a clinical and biomechanical analysis, J Bone Joint Surg Br, 70, 3, pp. 461-464, (1988)
[3]  
Hill J.M., McGuire M.H., Crosby L.A., Closed treatment of displaced middle-third fractures of the clavicle gives poor results, J Bone Joint Surg Br, 79, 4, pp. 537-539, (1997)
[4]  
Denard P.J., Koval K.J., Cantu R.V., Weinstein J.N., Management of midshaft clavicle fractures in adults, Am J Orthop, 34, 11, pp. 527-536, (2005)
[5]  
Wu C.C., Shih C.H., Chen W.J., Tai C.L., Treatment of clavicular aseptic nonunion: Comparison of plating and intramedullary nailing techniques, J Trauma, 45, 3, pp. 512-516, (1998)
[6]  
Proubasta I.R., Itarte J.P., Caceres E.P., Llusa M.P., Gil J.M., Planell J.A., Ginebra M.P., Biomechanical evaluation of fixation of clavicular fractures, J South Orthop Assoc, 11, 3, pp. 148-152, (2002)
[7]  
Harnroongroj T., Vanadurongwan V., Biomechanical aspects of plating osteosynthesis of transverse clavicular fracture with and without inferior cortical defect, Clin Biomech (Bristol, Avon), 11, 5, pp. 290-294, (1996)
[8]  
Iannotti M.R., Crosby L.A., Stafford P., Grayson G., Goulet R., Effects of plate location and selection on the stability of midshaft clavicle osteotomies: A biomechanical study, J Shoulder Elbow Surg, 11, 5, pp. 457-462, (2002)
[9]  
Capicotto P.N., Heiple K.G., Wilbur J.H., Midshaft clavicle nonunions treated with intramedullary Steinman pin fixation and onlay bone graft, J Orthop Trauma, 8, 2, pp. 88-93, (1994)
[10]  
Enneking T.J., Hartlief M.T., Fontijne W.P., Rushpin fixation for midshaft clavicular nonunions: Good results in 13/14 cases, Acta Orthop Scand, 70, 5, pp. 514-516, (1999)