Multi-channel electrical bioimpedance: A new noninvasive method to simultaneously measure cardiac and peripheral blood flow

被引:2
|
作者
Stanley Jr. A.W.H. [1 ,2 ,3 ]
Herald J.W. [2 ,3 ]
Athanasuleas C.L. [1 ,3 ,4 ]
Jacob S.C. [2 ,3 ]
Sims S.W. [2 ,3 ]
Bartolucci A.A. [5 ]
Tsoglin A.N. [6 ]
机构
[1] Kemp-Carraway Heart Institute, Birmingham, AL
[2] Cardiovascular Consultants of Alabama, Birmingham, AL
[3] Physicians Medical Center Carraway, Birmingham, AL
[4] Norwood Clinic, Birmingham, AL
[5] Department of Biostatistics, University of Alabama, Birmingham, AL
[6] Delta Segments Technology, Inc., Birmingham, AL
关键词
Bioimpedance; Multi-channel electrical bioimpedance; Noninvasive cardiac output; Noninvasive measurement of flow;
D O I
10.1007/s10877-007-9095-7
中图分类号
学科分类号
摘要
Objectives: We sought to assess the ability of a new multi-channel electrical bioimpedance (MEB) methodology to accurately measure both cardiac blood flow and peripheral limb blood flow. Background: Cardiac output is the primary determinant of peripheral blood flow; however, optimal regional tissue perfusion is ultimately dependent on the patency of the arterial conduits that transport that flow. A complete understanding of regional tissue perfusion requires knowledge of both cardiac and peripheral blood flow. Existing noninvasive devices do not simultaneously assess the cardiac and peripheral circulations. Methods: Cardiac blood flow (cardiac output) was measured by MEB in 30 healthy volunteers and was compared to a 2D-Echo Doppler cardiac output. Peripheral blood flow (regional ankle and arm flow) was measured by MEB in 15 healthy volunteers. The MEB ankle/arm flow ratio (AAI index) was then compared to a conventional ankle/brachial pressure ratio (ABI index). Results: There was good correlation between the mean cardiac index by MEB (3.08 l/min/m2) and by Echo Doppler (3.13 l/min/m2) and bias and precision was 0.051 (1.6%) and ±0.52 l/min/m2 (±17%), respectively. The close correlation was maintained for each measurement over a wide range of cardiac indices. There was good correlation between AAI and ABI measurements (p < 0.05) with a sensitivity of 100% and specificity of 100%. Conclusions: MEB methodology can precisely measure cardiac outputand peripheral limb flow in healthy volunteers. © Springer Science+Business Media B.V. 2007.
引用
收藏
页码:345 / 351
页数:6
相关论文
共 50 条
  • [1] MULTI-CHANNEL ELECTRICAL BIOIMPEDANCE: A NON-INVASIVE METHOD TO SIMULTANEOUSLY MEASURE CARDIAC OUTPUT AND INDIVIDUAL ARTERIAL LIMB FLOW IN PATIENTS WITH CARDIOVASCULAR DISEASE
    Stanley, Alfred W. H., Jr.
    Herald, Jeffery W.
    Athanasuleas, Constantine L.
    Jacob, Saji C.
    Bartolucci, Alfred A.
    Tsoglin, Alexander N.
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2009, 23 (04) : 243 - 251
  • [2] Multi-channel electrical bioimpedance: a non-invasive method to simultaneously measure cardiac output and individual arterial limb flow in patients with cardiovascular disease
    Alfred W. H. Stanley
    Jeffery W. Herald
    Constantine L. Athanasuleas
    Saji C. Jacob
    Alfred A. Bartolucci
    Alexander N. Tsoglin
    Journal of Clinical Monitoring and Computing, 2009, 23 : 243 - 251
  • [3] Assessment of the impact of acupuncture on peripheral blood flow with multi-channel photoplethysmography
    Yang, Cheng-Chan
    Zhuang, Wei-You
    Wu, Hsien-Tsai
    2014 IEEE INTERNATIONAL CONFERENCE ON ELECTRON DEVICES AND SOLID-STATE CIRCUITS (EDSSC), 2014,
  • [4] A New Method To Measure The Noise Figure Of Multi-Channel Radar Receiving System
    Li, Weimei
    Liu, Bo
    PROCEEDINGS OF THE 2017 2ND JOINT INTERNATIONAL INFORMATION TECHNOLOGY, MECHANICAL AND ELECTRONIC ENGINEERING CONFERENCE (JIMEC 2017), 2017, 62 : 252 - 255
  • [5] Multi-channel IP Electrical Method for Deep Exploration
    Liu Haifei
    Liu Jianxian
    Chen Rujun
    Guo Ronwen
    Tong Xiaozhong
    Peng Yanhua
    NEAR-SURFACE GEOPHYSICS AND ENVIRONMENT PROTECTION, 2012, : 162 - 167
  • [6] A New Method for Multi-channel Signal Compression
    Ding Lijun
    Feng Hao
    2011 INTERNATIONAL CONFERENCE ON ELECTRONICS, COMMUNICATIONS AND CONTROL (ICECC), 2011, : 67 - 70
  • [7] Research of multi-flow and multi-channel flow parts of the vortex expansion machines with the external peripheral channel
    Vaneev, S.
    Miroshnichenko, D.
    Meleychuk, S.
    Baga, V.
    XV INTERNATIONAL SCIENTIFIC AND ENGINEERING CONFERENCE HERMETIC SEALING, VIBRATION RELIABILITY AND ECOLOGICAL SAFETY OF PUMP AND COMPRESSOR MACHINERY (HERVICON+PUMPS-2017), 2017, 233
  • [8] A simplified noninvasive method to measure airway blood flow in humans
    Wanner, A
    Mendes, ES
    Atkins, ND
    JOURNAL OF APPLIED PHYSIOLOGY, 2006, 100 (05) : 1674 - 1678
  • [9] New multi-channel transcutaneous electrical stimulation technology for rehabilitation
    Keller, Thierry
    Lawrence, Marc
    Kuhn, Andreas
    Morari, Manfred
    2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 6191 - +
  • [10] A new method of multi-channel impedance rheography measurement
    Sha, H.
    Li, S.
    Cui, Y.
    Beijing Shengwu Yixue Gongcheng/Beijing Biomedical Engineering, 2001, 20 (02): : 126 - 128