HPLC column-switching technique for sample preparation and fluorescence determination of propranolol in urine using fused-core columns in both dimensions

被引:0
作者
Dalibor Šatínský
Lucie Havlíková
Petr Solich
机构
[1] Charles University,Department of Analytical Chemistry, Faculty of Pharmacy
来源
Analytical and Bioanalytical Chemistry | 2013年 / 405卷
关键词
Fused-core column; HPLC; Column switching; Large-volume sample; Propranolol; Urine;
D O I
暂无
中图分类号
学科分类号
摘要
A new and fast high-performance liquid chromatography (HPLC) column-switching method using fused-core columns in both dimensions for sample preconcentration and determination of propranolol in human urine has been developed. On-line sample pretreatment and propranolol preconcentration were performed on an Ascentis Express RP-C-18 guard column (5 × 4.6 mm), particle size, 2.7 μm, with mobile phase acetonitrile/water (5:95, v/v) at a flow rate of 2.0 mL min−1 and at a temperature of 50 °C. Valve switch from pretreatment column to analytical column was set at 4.0 min in a back-flush mode. Separation of propranolol from other endogenous urine compounds was achieved on the fused-core column Ascentis Express RP-Amide (100 × 4.6 mm), particle size, 2.7 μm, with mobile phase acetonitrile/water solution of 0.5 % triethylamine, pH adjusted to 4.5 by means of glacial acetic acid (25:75, v/v), at a flow rate of 1.0 mL min−1 and at a temperature of 50 °C. Fluorescence excitation/emission detection wavelengths were set at 229/338 nm. A volume of 1,500 μL of filtered urine sample solution was injected directly into the column-switching HPLC system. The total analysis time including on-line sample pretreatment was less than 8 min. The experimentally determined limit of detection of the method was found to be 0.015 ng mL−1.
引用
收藏
页码:6583 / 6587
页数:4
相关论文
共 61 条
[1]  
Sadílek P(2007)undefined Trac Trends Anal Chem 26 375-384
[2]  
Šatínský D(2007)undefined J Sep Sci 30 3104-3109
[3]  
Solich P(2013)undefined Talanta 103 221-227
[4]  
Cunliffe JM(2007)undefined J Sep Sci 30 3164-3173
[5]  
Maloney TD(2012)undefined Environ Sci Technol 46 6750-6758
[6]  
Chocholouš P(2009)undefined J Pharm Biomed Anal 50 966-976
[7]  
Vacková J(2003)undefined J Pharm Biomed Anal 33 699-709
[8]  
Šrámková I(2011)undefined Chin J Chromatogr 29 1205-1209
[9]  
Šatínský D(2011)undefined J Chromatogr B 879 615-626
[10]  
Solich P(2009)undefined J Chromatogr A 1216 190-197