Position and Angular Misalignment Analysis for Implantable Wireless Power Transfer System Based on Magnetic Resonance

被引:0
作者
Fei-Xiang Gong
Zhiqiang Wei
Haokun Chi
Bo Yin
Ying Sun
Yanping Cong
Mingui Sun
机构
[1] Ocean University of China,School of Information Science and Engineering
[2] University of Pittsburgh,Department of Neurological Surgery
[3] Hebei University of Technology,School of Electrical Engineering
来源
Journal of Medical and Biological Engineering | 2017年 / 37卷
关键词
Position and angular misalignment; Magnetic resonance; Wireless power transfer; Transfer efficiency;
D O I
暂无
中图分类号
学科分类号
摘要
As a widely-acknowledged truth, the interval of human tissue, the breathing activity, and the body motion work together to change the relative position of the antenna. This knowledge, with detailed examination, is going to bring significant change to the transmission efficiency in the implementation of human wireless power transfer. This paper focused on the relationship of the transfer efficiency with lateral distance, vertical distance and angular (around the center and along the edge) misalignment. A new formula, which serves to connect the geometrical parameters both with the self-resonance frequency and with the Q-factor of the printed circuit board antenna, was given. Furthermore, the paper gave a significant analysis of the efficiency with lateral distance change, vertical distance change and two kinds of angular misalignments using finite element method. The experimental results showed that in the situation of vertical distance change and angular (along the edge) misalignment, the transmission efficiency drops rapidly with the displacement of antenna. However, in the situation of lateral distance change and angular (along the edge) misalignment, there were a high-efficiency distance—the lateral misalignment ≤40% of the length of the width in the lateral misalignment; and a high-efficiency angle—angular around the center misalignment ≤30º in the angular misalignment. Within the ranges, a high-efficiency space (HES) is formed in the implantable WPT system utility, and the transmission efficiency drops rapidly when the antenna is beyond the HES. This paper can provide a practical application to the antenna design and specific implementation in human implantable WPT system.
引用
收藏
页码:602 / 611
页数:9
相关论文
共 58 条
  • [11] Kiani M(2011)A design method of magnetically resonanting wireless power delivery systems for bio-implantable devices IEEE Transactions on Magnetics 47 3833-3836
  • [12] Ghovanloo M(2013)High-efficiency wireless power transfer for biomedical implants by optimal resonant load transformation IEEE Transactions on Circuits and Systems I: Regular Papers 60 867-874
  • [13] Wen-Zhen FU(2013)On the design of efficient multi-coil telemetry system for biomedical implants IEEE Transactions on Biomedical Circuits and Systems 7 11-23
  • [14] Zhang B(2012)Lateral and angular misalignments analysis of a new PCB circular spiral resonant wireless charger IEEE Transactions on Magnetics 48 4522-4525
  • [15] Qiu DY(2014)Field distribution models of spiral coil for misalignment analysis in wireless power transfer systems IEEE Transactions on Microwave Theory and Techniques 62 920-930
  • [16] Wang W(2011)Wireless power transfer in loosely coupled links: coil misalignment model IEEE Transactions on Magnetics 47 416-430
  • [17] Karalis A(undefined)undefined undefined undefined undefined-undefined
  • [18] Joannopoulos JD(undefined)undefined undefined undefined undefined-undefined
  • [19] Soljačić M(undefined)undefined undefined undefined undefined-undefined
  • [20] Li X(undefined)undefined undefined undefined undefined-undefined