A simplified 4-parameter model of volumetric capnograms improves calculations of airway dead space and slope of Phase III

被引:0
作者
Gabriel C. Motta-Ribeiro
Marcos F. Vidal Melo
Frederico C. Jandre
机构
[1] COPPE,Biomedical Engineering Programme
[2] Universidade Federal Do Rio de Janeiro,Department of Anesthesia, Critical Care and Pain Medicine
[3] Massachusetts General Hospital and Harvard Medical School,undefined
来源
Journal of Clinical Monitoring and Computing | 2020年 / 34卷
关键词
CO; Volumetric capnogram; Dead space; Phase III slope; Mathematical modeling;
D O I
暂无
中图分类号
学科分类号
摘要
To evaluate a compact and easily interpretable 4-parameter model describing the shape of the volumetric capnogram, and the resulting estimates of anatomical dead space (VDAW) and Phase III (alveolar plateau) slope (SIII). Data from of 8 mildly-endotoxemic pre-acute respiratory distress syndrome sheep were fitted to the proposed 4-parameter model (4p) and a previously established 7-parameter model (7p). Root mean square error (RMSE) and Akaike information criterion (AIC), as well as VDAW and SIII derived from each model were compared. Confidence intervals for model’s parameters, VDAW and SIII were estimated with a jackknife approach. RMSE values were similar (4p: 1.13 ± 0.01 mmHg vs 7p: 1.14 ± 0.01 mmHg) in the 791 breath cycles tested. However, the 7p overfitted the curve and had worse AIC in more than 50% of the cycles (p < 0.001). The large number of degrees of freedom also resulted in larger between-animal range of confidence intervals for 7p (VDAW: from 6.1 10–12 to 34 ml, SIII: from 9.53 10–7 to 1.80 mmHg/ml) as compared to 4p (VDAW: from 0.019 to 0.15 ml, SIII: from 3.9 10–4 to 0.011 mmHg/ml). Mean differences between VDAW (2.1 ± 0.04 ml) and SIII (0.047 ± 0.004 mmHg/ml) from 7 and 4p were significant (p < 0.001), but within the observed cycle-by-cycle variability. The proposed 4-parameter model of the volumetric capnogram improves data fitting and estimation of VDAW and SIII as compared to the 7-parameter model of reference. These advantages support the use of the 4-parameter model in future research and clinical applications.
引用
收藏
页码:1265 / 1274
页数:9
相关论文
共 137 条
[1]  
Tusman G(2011)Capnography reflects ventilation/perfusion distribution in a model of acute lung injury Acta Anaesthesiol Scand 55 597-606
[2]  
Suarez-Sipmann F(2014)Volumetric capnography: the time has come Curr Opin Crit Care 20 333-339
[3]  
Bohm HS(2018)Volumetric but not time capnography detects ventilation/perfusion mismatch in injured rabbit lung Front Physiol 9 1805-834
[4]  
Borges BJ(2019)Clinical use of volumetric capnography in mechanically ventilated patients J Clin Monit Comput 98 828-167
[5]  
Hedenstierna G(2007)Systematic errors and susceptibility to noise of four methods for calculating anatomical dead space from the CO2 expirogram Br J Anaesth 32 161-206
[6]  
Suarez-Sipmann F(1999)An algebraic solution to dead space determination according to Fowler’s graphical method Comput Biomed Res 23 197-190
[7]  
Bohm SH(2009)Model fitting of volumetric capnograms improves calculations of airway dead space and slope of Phase III J Clin Monit Comput 26 183-288
[8]  
Tusman G(2012)States of low pulmonary blood flow can be detected non-invasively at the bedside measuring alveolar dead space J Clin Monit Comput 27 281-238
[9]  
Tolnai J(2013)Reference values for volumetric capnography-derived non-invasive parameters in healthy individuals J Clin Monit Comput 45 231-286
[10]  
Fodor GH(2018)Dead space analysis at different levels of positive end-expiratory pressure in acute respiratory distress syndrome patients J Crit Care 12 283-88