Effect of x-ray tube potential on image quality and patient dose for lumbar spine computed radiography examinations

被引:8
|
作者
Brindhaban A. [1 ]
Al Khalifah K. [1 ]
Al Wathiqi G. [2 ]
Al Ostath H. [3 ]
机构
[1] Department of Radiologic Sciences, Kuwait University, Sulaibikhat 90805
[2] Mubarak Al Kabir Hospital, Ministry of Health
[3] Amiri Hospital, Ministry of Health
来源
Australasian Physics & Engineering Sciences in Medicine | 2005年 / 28卷 / 4期
关键词
Image quality; Lumbar spine; Patient dose; X-ray tube potential;
D O I
10.1007/BF03178721
中图分类号
学科分类号
摘要
A study was conducted to investigate the effect of increasing x-ray tube voltage on patient dose and image quality during computed radiography (CR) examination of the lumbar spine. Images of a pelvis - lumbar spine phantom were obtained at three different tube voltages using two CR systems. The images were evaluated by image scores (IS) using the image quality criteria proposed by the Commission of the European Communities (CEC), by measuring the signal-to-noise ratio and the contrast-to-noise ratio. The entrance surface dose (ESD) to the phantom was also measured using ionisation chambers and the resulting gonad dose (G) and effective dose (E) were calculated. The CEC recommended tube voltages can be exceeded with CR systems for lumbar spine radiography. The ESD, G and E decreased by 25% - 50% with both CR systems when tube voltage was increased. Although the decrease in IS was statistically significant (P<0.05) the images obtained with the 3 tube voltages were diagnostically acceptable. The balance between image quality and patient dose can be optimised by using higher x-ray tube voltages for lumbar spine CR examinations. Copyright © 2005 ACPSEM/EA.
引用
收藏
页码:216 / 222
页数:6
相关论文
共 50 条
  • [1] Comparison of image quality and patient dose for chest x-ray examinations on conventional and low cost computed radiography systems
    Muhogora, W. E.
    Trianni, A.
    Toso, F.
    Devetti, A.
    Padovani, R.
    Msaki, P.
    Kazema, R.
    RADIOGRAPHY, 2012, 18 (04) : 275 - 278
  • [2] PATIENT DOSE AND IMAGE QUALITY IN 5 STANDARD X-RAY EXAMINATIONS
    HAVUKAINEN, R
    PIRINEN, M
    MEDICAL PHYSICS, 1993, 20 (03) : 813 - 817
  • [3] Effect of varying X-ray tube voltage and additional filtration on image quality and patient dose in digital radiography system
    Aksit, E.
    Yalcin, A.
    Olgar, T.
    APPLIED RADIATION AND ISOTOPES, 2023, 199
  • [4] Effect of Tube Setting on Image Quality in Industrial X-ray Computed Tomography
    Sun, Zhaoying
    Li, Lei
    Xi, Xiaoqi
    Han, Yu
    Yan, Bin
    Rong, Lihui
    Zhou, Sangsang
    THREE-DIMENSIONAL IMAGE ACQUISITION AND DISPLAY TECHNOLOGY AND APPLICATIONS, 2018, 10845
  • [5] A baseline study of entrance dose and image quality for lumbar spine radiography in Calabar, Nigeria
    Egbe, N. O.
    Chiaghanam, N. O.
    Azogor, W. E.
    Inyang, S. O.
    RADIOGRAPHY, 2009, 15 (04) : 306 - 312
  • [6] Radiation dose to patients from X-ray radiographic examinations using computed radiography imaging system
    Sharma, Reena
    Sharma, Sunil Dutt
    Pawar, Shramika
    Chaubey, Ajay
    Kantharia, S.
    Babu, D. A. R.
    JOURNAL OF MEDICAL PHYSICS, 2015, 40 (01) : 29 - 37
  • [7] Optimizing abdominal CT dose and image quality with respect to x-ray tube voltage
    Huda, W
    Ogden, KM
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 499 - 507
  • [8] AP versus PA positioning in lumbar spine computed radiography: Image quality and individual organ doses
    Davey, Enda
    England, Andrew
    RADIOGRAPHY, 2015, 21 (02) : 188 - 196
  • [9] The effect of patient shield position on gonad dose during lumbar spine radiography
    Clancy, Conor L.
    O'Reilly, Geraldine
    Brennan, Patrick C.
    McEntee, Mark F.
    RADIOGRAPHY, 2010, 16 (02) : 131 - 135
  • [10] Radiation dose reduction in computed skeletal radiography - Effect on image quality
    Jonsson, A
    Herrlin, K
    Jonsson, K
    Lundin, B
    Sanfridsson, J
    Pettersson, H
    ACTA RADIOLOGICA, 1996, 37 (02) : 128 - 133