Clinical value of stereoscopic three-dimensional echocardiography in assessment of atrial septal defects: Feasibility and efficiency

被引:0
作者
Jing Wang
Xinfang Wang
Mingxing Xie
Lin He
Qing Lv
Lei Wang
机构
[1] Huazhong University of Science and Technology,Department of Ultrasonography, Union Hospital
[2] Hubei Provincial Key Laboratory of Molecular Imaging,undefined
来源
Journal of Huazhong University of Science and Technology [Medical Sciences] | 2009年 / 29卷
关键词
echocardiography; stereoscopic vision; real time; three-dimension; atrial septal defect; occlusion;
D O I
暂无
中图分类号
学科分类号
摘要
Stereoscopic three-dimensional echocardiography(S-3DE) is a novel displaying technology based on real-time 3-dimensional echocardiography (RT-3DE). Our study was to evaluate the feasibility and efficiency of S-3DE in the diagnosis of atrial septal defect (ASD) and its use in the guidance for transcatheter ASD occlusion. Twelve patients with secundum ASD underwent RT-3DE examination and 9 of the 12 were subjected to transcatheter closure of ASD. Stereoscopic vision was generated with a high-performance volume renderer with red-green stereoscopic glasses. S-3DE was compared with standard RT-3D display for the assessment of the shape, size, and the surrounding tissues of ASD and for the guidance of ASD occlusion. The appearance rate of coronary sinus and the mean formation time of the IVC, SVC were compared. Our results showed that S-3DE could measure the diameter of ASD accurately and there was no significant difference in the measurements between S-3DE and standard 3D display (2.89±0.73 cm vs 2.85±0.72 cm, P>0.05; r=0.96, P<0.05). The appearance of coronary sinus for S-3DE was higher as compared with the standard 3D display (93.3% vs 100%). The mean time of the IVC, SVC for S-3DE monitor was slightly shorter than that of the standard 3D display (11.0±3.8 s vs 10.3±3.6 s, P>0.05). The mean completion time of interventional procedure was shortened with S-3DE display as compared with standard 3D display (17.3±3.1 min vs 23.0±3.9 min, P<0.05). Stereoscopic three-dimensional echocardiography could improve the visualization of three-dimensional echocardiography, facilitate the identification of the adjacent structures, decrease the time required for interventional manipulation. It may be a feasible, safe, and efficient tool for guiding transcatheter septal occlusion or the surgical interventions.
引用
收藏
页码:791 / 794
页数:3
相关论文
共 42 条
  • [1] Correale M.(2008)Real-time three-dimensional echocardiography: an update Eur J Intern Med 19 241-248
  • [2] Ieva R.(2008)Real-time transesophageal three-dimensional echocardiography for guidance of percutaneous cardiac interventions:first experience Clin Res Cardiol 97 565-574
  • [3] Di Biase M.(2009)Real-time three-dimensional echocardiographic assessment of left ventricular remodeling index in patients with hypertensive heart disease and coronary artery disease J Huazhong Univ Sci Technolog [Med Sci] 29 122-126
  • [4] Balzer J.(2000)Transcatheter closure of multiple atrial septal defects. Initial results and value of two- and three-dimensional transoesophageal echocardiography Eur Heart J 21 941-947
  • [5] Kühl H.(2006)Real timethree-dimensional transthoracic echocardiography for guiding Amplatzer septal occluder device deployment in patients with atrial septal defect Echocardiography 23 763-770
  • [6] Rassaf T.(2009)Optimization of ASD assessment using real time three-dimensional transesophageal echocardiography Echocardiography 26 233-235
  • [7] Chen M.(2001)Transcatheter closure of atrial septal defects and patent foramen ovale under intracardiac echocardiographic guidance: feasibility and comparison with transesophageal echocardiography Catheter Cardiovasc Interv 52 194-199
  • [8] Wang J.(2004)Intracardiac echocardiographic guidance during transcatheter device closure of atrial septal defect and patent foramen ovale Mayo Clin Proc 79 24-34
  • [9] Xie M.(2008)Usefulness of intraoperative real-time 3D transesophageal echocardiography in cardiac surgery J Card Surg 23 784-786
  • [10] Cao Q.(2004)Real-time 3-dimensional echocardiography in assessing atrial and ventricular septal defects: an echocardiographic-surgical correlative study Am Heart J 148 1091-1095