Tendon–bone contact pressure and biomechanical evaluation of a modified suture-bridge technique for rotator cuff repair

被引:0
作者
Mike H. Baums
Michael Geyer
Meike Büschken
Gottfried H. Buchhorn
Gunter Spahn
Hans-Michael Klinger
机构
[1] University of Göttingen Medical Centre (UMG),Department of Orthopaedic Surgery
[2] Georg-August-University,Department of Orthopaedic Surgery
[3] St. Vinzenz Klinik Pfronten,undefined
[4] Clinic of Traumatology and Orthopaedic Surgery,undefined
来源
Knee Surgery, Sports Traumatology, Arthroscopy | 2010年 / 18卷
关键词
Rotator cuff repair; Contact pressure; Suture-bridge; Sheep shoulder;
D O I
暂无
中图分类号
学科分类号
摘要
The aim of the study was to evaluate the time-zero mechanical and footprint properties of a suture-bridge technique for rotator cuff repair in an animal model. Thirty fresh-frozen sheep shoulders were randomly assigned among three investigation groups: (1) cyclic loading, (2) load-to-failure testing, and (3) tendon–bone interface contact pressure measurement. Shoulders were cyclically loaded from 10 to 180 N and displacement to gap formation of 5- and 10-mm at the repair site. Cycles to failure were determined. Additionally, the ultimate tensile strength and stiffness were verified along with the mode of failure. The average contact pressure and pressure pattern were investigated using a pressure-sensitive film system. All of the specimens resisted against 3,000 cycles and none of them reached a gap formation of 10 mm. The number of cycles to 5-mm gap formation was 2,884.5 ± 96.8 cycles. The ultimate tensile strength was 565.8 ± 17.8 N and stiffness was 173.7 ± 9.9 N/mm. The entire specimen presented a unique mode of failure as it is well known in using high strength sutures by pulling them through the tendon. We observed a mean contact pressure of 1.19 ± 0.03 MPa, applied on the footprint area. The fundamental results of our study support the use of a suture-bridge technique for optimising the conditions of the healing biology of a reconstructed rotator cuff tendon. Nevertheless, an individual estimation has to be done if using the suture-bridge technique clinically. Further investigation is necessary to evaluate the cell biological healing process in order to achieve further sufficient advancements in rotator cuff repair.
引用
收藏
页码:992 / 998
页数:6
相关论文
共 156 条
[1]  
Apreleva M(2002)Rotator cuff tears: the effect of the reconstruction method on three-dimensional repair site area Arthroscopy 18 519-526
[2]  
Özbaydar M(2008)Biomechanical characteristics of single-row repair in comparison to double-row repair with consideration of the suture configuration and suture material Knee Surg Sports Traumatol Arthrosc 16 1052-1060
[3]  
Fitzgibbons PG(2006)Cuff integrity after arthroscopic versus open rotator cuff repair: a prospective study J Shoulder Elbow Surg 15 290-299
[4]  
Warner JJP(2008)Influence of suture material on the biomechanical behaviour of suture-tendon specimens Am J Sports Med 36 907-912
[5]  
Baums MH(2007)A novel double-row rotator cuff repair exceeds strengths of conventional repairs Clin Orthop Rel Res 461 106-113
[6]  
Buchhorn GH(1997)Cyclic loading of anchor-based rotator cuff repairs: confirmation of the tension overload phenomenon and comparison of suture anchor fixation with transosseous fixation Arthroscopy 13 720-724
[7]  
Spahn G(1997)Cyclic loading of transosseous rotator cuff repairs: tension overload as a possible cause of failure Arthroscopy 13 172-176
[8]  
Poppendieck B(2008)A biomechanical comparison of 2 technical variations of double-row rotator cuff fixation Am J Sports Med 36 901-906
[9]  
Schultz W(2007)Can a double-row anchorage technique improve tendon healing in arthroscopic rotator cuff repair? A prospective, nonrandomized, comparative study of double-row and single-row anchorage techniques with computed tomography arthrography tendon healing assessment Am J Sports Med 35 1247-1253
[10]  
Klinger HM(2007)The Roman Bridge: a “double pulley-suture bridges” technique for rotator cuff repair BMC Musculoskelet Disord 8 123-129