Short uncemented stems allow greater femoral flexibility and may reduce peri-prosthetic fracture risk: a dry bone and cadaveric study

被引:12
作者
Jones C. [1 ]
Aqil A. [1 ]
Clarke S. [1 ]
Cobb J.P. [1 ]
机构
[1] MSK Lab, 7th Floor, Lab Block, Department of Surgery and Cancer, Imperial College London, Charing Cross Hospital, Fulham Palace Road, London
关键词
Fracture; Mechanical testing; Short stem; Total hip arthroplasty;
D O I
10.1007/s10195-015-0343-1
中图分类号
学科分类号
摘要
Background: Short femoral stems for uncemented total hip arthroplasty have been introduced as a safe alternative to traditional longer stem designs. However, there has been little biomechanical examination of the effects of stem length on complications of surgery. This study aims to examine the effect of femoral stem length on torsional resistance to peri-prosthetic fracture. Materials and methods: We tested 16 synthetic and two paired cadaveric femora. Specimens were implanted and then rapidly rotated until fracture to simulate internal rotation on a planted foot, as might occur during stumbling. 3D planning software and custom-printed 3D cutting guides were used to enhance the accuracy and consistency of our stem insertion technique. Results: Synthetic femora implanted with short stems fractured at a significantly higher torque (27.1 vs. 24.2 Nm, p = 0.03) and angle (30.3° vs. 22.3°, p = 0.002) than those implanted with long stems. Fracture patterns of the two groups were different, but showed remarkable consistency within each group. These characteristic fracture patterns were closely replicated in the pair of cadaveric femora. Conclusions: This new short-stemmed press-fit femoral component allows more femoral flexibility and confers a higher resistance to peri-prosthetic fracture from torsional forces than long stems. © 2015, The Author(s).
引用
收藏
页码:229 / 235
页数:6
相关论文
共 21 条
  • [1] Lombardi A.V., Berend K.R., Ng V.Y., Stubby stems: good things come in small packages, Orthopedics, 34, pp. e464-e466, (2011)
  • [2] McElroy M.J., Johnson A.J., Mont M.A., Bonutti P.M., Short and standard stem prostheses are both viable options for minimally invasive total hip arthroplasty, Bull NYU Hosp Jt Dis, 69, pp. S68-S76, (2011)
  • [3] Molli R.G., Lombardi A.V., Berend K.R., Adams J.B., Sneller M.A., A short tapered stem reduces intraoperative complications in primary total hip arthroplasty, Clin Orthop Relat Res, 470, pp. 450-461, (2012)
  • [4] McLaughlin J.R., Lee K.R., Cementless total hip replacement using second-generation components: a 12- to 16-year follow-up, J Bone Joint Surg Br, 92, pp. 1636-1641, (2010)
  • [5] Sariali E., Mouttet A., Mordasini P., Catonne Y., High 10-year survival rate with an anatomic cementless stem (SPS), Clin Orthop Relat Res, 470, pp. 1941-1949, (2012)
  • [6] Berry D.J., Epidemiology: hip and knee, Orthop Clin North Am, 30, pp. 183-190, (1999)
  • [7] Meek R.M., Norwood T., Smith R., Brenkel I.J., Howie C.R., The risk of peri-prosthetic fracture after primary and revision total hip and knee replacement, J Bone Joint Surg Br, 93, pp. 96-101, (2011)
  • [8] Lindahl H., Oden A., Garellick G., Malchau H., The excess mortality due to periprosthetic femur fracture. A study from the Swedish national hip arthroplasty register, Bone, 40, pp. 1294-1298, (2007)
  • [9] Schmidt A.H., Kyle R.F., Periprosthetic fractures of the femur, Orthop Clin North Am, 33, pp. 143-152, (2002)
  • [10] Jakubowitz E., Seeger J.B., Lee C., Heisel C., Kretzer J.P., Thomsen M.N., Do short-stemmed-prostheses induce periprosthetic fractures earlier than standard hip stems? A biomechanical ex vivo study of two different stem designs, Arch Orthop Trauma Surg, 129, pp. 849-855, (2009)