Coronary artery calcium scoring with photon-counting CT: first in vivo human experience

被引:0
作者
Rolf Symons
Veit Sandfort
Marissa Mallek
Stefan Ulzheimer
Amir Pourmorteza
机构
[1] Radiology and Imaging Sciences - National Institutes of Health Clinical Center,Department of Imaging & Pathology
[2] University Hospitals Leuven,Department of Radiology and Imaging Sciences
[3] Siemens Healthcare GmbH,Department of Radiology and Imaging Sciences
[4] Emory University School of Medicine,undefined
[5] Winship Cancer Institute of Emory University,undefined
来源
The International Journal of Cardiovascular Imaging | 2019年 / 35卷
关键词
Coronary artery calcium score; Photon-counting CT; Radiation dose reduction;
D O I
暂无
中图分类号
学科分类号
摘要
To evaluate the performance of photon-counting detector (PCD) computed tomography (CT) for coronary artery calcium (CAC) score imaging at standard and reduced radiation doses compared to conventional energy-integrating detector (EID) CT. A dedicated cardiac CT phantom, ten ex vivo human hearts, and ten asymptomatic volunteers underwent matched EID and PCD CT scans at different dose settings without ECG gating. CAC score, contrast, and contrast-to-noise ratio (CNR) were calculated in the cardiac CT phantom. CAC score accuracy and reproducibility was assessed in the ex vivo hearts. Standard radiation dose (120 kVp, reference mAs = 80) in vivo CAC scans were compared against dose-reduced CAC scans (75% dose reduction; reference mAs = 20) for image quality and CAC score reproducibility. Interstudy agreement was assessed by using intraclass correlation (ICC), linear regression, and Bland–Altman analysis with 95% confidence interval limits of agreement (LOA). Calcium-soft tissue contrast and CNR were significantly higher for the PCD CAC scans in the cardiac CT phantom (all P < 0.01). Ex vivo hearts: CAC score reproducibility was significantly higher for the PCD scans at the lowest dose setting (50 mAs) (P = 0.002); score accuracy was similar for both detector systems at all dose settings. In vivo scans: the agreement between standard dose and low dose CAC score was significantly better for the PCD than for the EID with narrower LOA in Bland–Altman analysis, linear regression slopes closer to 1 (0.96 vs. 0.84), and higher ICC values (0.98 vs. 0.93, respectively). Phantom and in vivo human studies showed PCD may significantly improve CAC score image quality and/or reduce CAC score radiation dose while maintaining diagnostic image quality.
引用
收藏
页码:733 / 739
页数:6
相关论文
共 50 条
  • [31] Photon-Counting Detector CT for Femoral Stent Imaging in an Extracorporeally Perfused Human Cadaveric Model
    Hartung, Viktor
    Gruschwitz, Philipp
    Huflage, Henner
    Augustin, Anne Marie
    Kleefeldt, Florian
    Peter, Dominik
    Lichthardt, Sven
    Erguen, Sueleyman
    Bley, Thorsten Alexander
    Grunz, Jan-Peter
    Petritsch, Bernhard
    INVESTIGATIVE RADIOLOGY, 2024, 59 (04) : 320 - 327
  • [32] Coronary artery calcium scoring in the age of CT angiography: What is its role?
    Paolo Raggi
    Akbar Khan
    Chesnal Arepali
    Arthur E. Stillman
    Current Atherosclerosis Reports, 2008, 10 : 438 - 443
  • [33] Photon-counting CT systems: A technical review of current clinical possibilities
    Greffier, Joel
    Viry, Anais
    Robert, Antoine
    Khorsi, Mouad
    Si-Mohamed, Salim
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2025, 106 (02) : 53 - 59
  • [34] A methodology for incorporating a photon-counting CT system into routine clinical use
    Bache, Steven T.
    Samei, Ehsan
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (08):
  • [35] Dental imaging in clinical photon-counting CT at a quarter of DVT dose
    Sawall, Stefan
    Maier, Joscha
    Sen, Sinan
    Gehrig, Holger
    Kim, Ti -Sun
    Schlemmer, Heinz -Peter
    Schoenberg, Stefan O.
    Kachelriess, Marc
    Ruetters, Maurice
    JOURNAL OF DENTISTRY, 2024, 142
  • [36] Imaging the pancreas with photon-counting CT - A review of normal pancreatic anatomy
    Brandt, Erik G. S.
    Mueller, Christoph F.
    Thomsen, Henrik
    Rodell, Anders B.
    Ibragimov, Bulat
    Andersen, Michael B.
    EUROPEAN JOURNAL OF RADIOLOGY, 2024, 181
  • [37] Photon-counting normalized metal artifact reduction (NMAR) in diagnostic CT
    Byl, Achim
    Klein, Laura
    Sawall, Stefan
    Heinze, Sarah
    Schlemmer, Heinz-Peter
    Kachelriess, Marc
    MEDICAL PHYSICS, 2021, 48 (07) : 3572 - 3582
  • [38] A systematic assessment and optimization of photon-counting CT for lung density quantifications
    Sotoudeh-Paima, Saman
    Segars, W. Paul
    Ghosh, Dhrubajyoti
    Luo, Sheng
    Samei, Ehsan
    Abadi, Ehsan
    MEDICAL PHYSICS, 2024, 51 (04) : 2893 - 2904
  • [39] Feasibility of the automated detection of colorectal polyps in photon-counting CT colonography
    Nappi, Janne J.
    Hironaka, Toru
    Wu, Dufan
    Yoshida, Stephen R.
    Gupta, Rajiv
    Tachibana, Rie
    Taguchi, Katsuyuki
    Yoshida, Hiroyuki
    MEDICAL IMAGING 2023, 2023, 12469
  • [40] Photon-counting CT: technical features and clinical impact on abdominal imaging
    Onishi, Hiromitsu
    Tsuboyama, Takahiro
    Nakamoto, Atsushi
    Ota, Takashi
    Fukui, Hideyuki
    Tatsumi, Mitsuaki
    Honda, Toru
    Kiso, Kengo
    Matsumoto, Shohei
    Kaketaka, Koki
    Enchi, Yukihiro
    Kawabata, Shuichi
    Nakasone, Shinya
    Tomiyama, Noriyuki
    ABDOMINAL RADIOLOGY, 2024, 49 (12) : 4383 - 4399